13
while raising the stopper, and the film magazine will come out.
•
Load a new full film magazine into the camera chamber.
•
Load an empty receiver.
•
Close the front door of the camera chamber.
•
Turn the CAMERA EVAC switch to EVAC. Usually after about 10 minutes. EVAC (green) lamp of
VACCUM STATE lights up.
•
Loading of new film
1) Take out exposed film and the light shielding plate
2) Place film in the magazine.
3) After loading the films, don’t forget to insert the light shielding plate
•
Open the desiccator’s door and introduce the full film magazine.
•
Turn the knob to EVAC.
•
Reset film number to 50 on the CRT using the keypad. Listen for the shielding plate to properly feed
through into the film receiver.
•
Change the FILM NUMBER to 0 using the keypad.
20. Final check procedure
•
Turn off all lights (room light, CRT monitor, panel light).
•
Complete the log book.
•
Fill final check list.