Excitation Setting Adjustment 7-34
2
7.13 Excitation Setting Adjustment
Press Excitation V update (nozzle test) .
(1) Overview
¡
Excitation V-ref. are from 0 to 19.
¡
To maintain good printing quality, it is necessary to input an optimum excitation
V-ref..
¡
Conduct a nozzle property test, and set a center value in a range capable of
good printing as the optimum excitation V-ref..
(ex.) In the case where the good printing is performed when the excitation
V-ref. is 5 to 15 in the nozzle property test, the optimum excitation
V-ref. is 10 at the center.
¡
An ambient temperature when updating the excitation V-ref. is stored as a
reference temperature. If the difference between the ambient temperature and
the reference ambient temperature exceeds a certain value, a warning
"Excitation V-ref. Review" arises. Update the excitation V-ref. in that case.
(2) Operation
Maintenance
Auxiliary
function
Show fault
Menu
Operation
management
Environment
setup menu
The "Maintenance" screen then opens.
Maintenance menu
2011.07.07 12:45
Environment
setup menu
Operation
management
Test Print
Excitation V update
(Nozzle test)
View software
version
View alarm
history
Circulation control
Password
protection
Comm. monitor
Back
Manual
Start
up
Shut
down
Comm
On/Off
Status: Standby Com=0
Maintenance menu
2011.07.07 12:45
Environment
setup menu
Operation
management
Test Print
Excitation V update
(Nozzle test)
View alarm
history
Circulation control
Manual
Start
up
Shut
down
Comm
On/Off
Status: Standby Com=0
1
Press Maintenance from the menu.
Summary of Contents for PXR-D