(1) Turn on the earth leakage breaker on the back of the main unit.

(2) Turn on the power switch on the right side of the main unit to start the 
device.

(3) When the unit is turned on the Evacuation process will start 
automatically and the EVAC LED will blink BLUE.

If you want to load a sample, press the EVAC/AIR button. The AIR LED will 
then blink WHITE. Once the specimen chamber is at air, the AIR LED will be 
stop blinking and remain WHITE. 

(4) Turn on the PC. Start the TM4000/TM4000Plus application by double-
clicking the TM4000 programme shortcut. The TM4000 start screen will be 
displayed until the Graphical User Interface (GUI) is loaded.

Bulk specimen (Conductive/Non-conductive)

Water / Oil containing specimen

(5) Put the conductive tape on the specimen stub and attach 
a specimen on it.

(5) When observing water containing specimens such as 
biological and botanical specimens, food items and so on, 
use a paste to stick a specimen on the stub.

(6) Attach the specimen stub to the specimen holder, and adjust its height using a height gauge so that the 
gap width becomes about 1 mm.

TM4000/TM4000Plus

Simple Manual

Front of TM4000

1. Starting The Instrument

2. Preparing Specimen

Page 1