Technical Specification
WARNING !
1. Heart disease patients SHOULD NOT use this device except under advice of a medical doctor.
2. DO NOT place the device in the mouth, neck and eyelid region.
3. This device is NOT suitable for persons implanted with cardiac pacemakers, and those
monitored by electronic equipment (e.g. ECG monitors and ECG alarms)
4. Patients with acute diseases and infectious diseases may use this device only under a medical
doctor's advice.
5. Patients with severe skin diseases or sensitive skin, or those in a state of bleeding SHOULD
NOT use this device.
6. Operation in close proximity (e.g. 1m) to a short wave therapy equipment may produce
instability in the device output.
1. Pulse Amplitude
:
Fixed 235 µA peak into 1,000 load
2. Pulse Rate
:
Fixed 16 Seconds/Cycle
3. Pulse Width
:
Fixed 800µs Voltage 4-16V peak into 1,000 load
4. Wave Form
:
Asymmetrical Bi-Phasic Square Pulse
5. Duration of Use
:
30 Minutes.
6. Motor Wave Form
:
2 Seconds On, 0.5 Seconds Off
7. Power Source
:
1.5 Volt Battery (size:AAA)
8. Operation Environment
:
10°C ~ +40°C , 30 ~ 85% RH
9. Storage Environment
:
-20°C ~ +50°C , 10 ~ 95% RH
10. Transportation Environment
:
-10°C ~ +50°C , 35 ~ 85% RH
( All electrical specification are ±20% 1,000 load )
* Design and specifications are subject to change without notice.
Note
:
This device complies with the electromagnetic compatibility requirement of EN 60601-1-2 as
specified in EEC Directive 93/42/EEC.