71
EN
GLISH
Temperature too low
Display:
„Lt“ & „Thermometer symbol“
The operating temperature is too low (below the
required temperature range from 10°C - 40°C
(50°F - 104°F)). The error is a warning that a
continuation under these conditions might lead
to wrong blood glucose readings.
Solution:
Take the meter to a location with ap-
propriate working temperature and wait for the
next measurement until the error does not oc-
cur again.
System error
Display:
„FAL“
Solution:
Re-insert the check strip (with writing
to top). If „FAL“ still occurs, please contact your
local supplier.
Error 002
Display:
„ERR 002“
Solution:
Device was not set by gateway. When
inserting batteries again, please be close to a
gateway. Device will anyhow try to connect with
gateway in certain interval.