background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

3

Button Push” or with a delayed start. The blue LED blinks every 
4 seconds when an alarm is tripped (unless the Show LED 
option is disabled). The blue and red LEDs will blink once when 
you press the magnetic switch to wake up the logger or will 
blink four times when you press the switch to start or stop 
logging. If you select Page Logger LED in the HOBOmobile app, 
the red LED will be illuminated for 5 seconds. 

Mounting Hole:

 Use the hole on the mounting end of the 

logger to attach it to a cable or cord for deployment (see 

Deploying and Mounting the Logger

). 

Switch:

 This magnetic switch on the mounting end of the logger 

works in conjunction with the reed switch. Press this switch for 3 
seconds to start or stop the logger when it is configured to start 
or stop “On Button Push” (see 

Configuring the Logger

). Press this 

switch for 1 second to wake up the logger (if configured with 
Bluetooth Always Off as described in 

Configuring the Logger

). 

Installing the pH Electrode 

The logger is shipped with the pH electrode uninstalled and the 
glass pH sensor placed in a small bottle filled with storage 
solution as shown below. 

Important:

 

The pH sensor must be kept in storage solution 

when not being calibrated or deployed in water

. The hydration 

layer surrounding the sensor starts to break down after 10 
minutes if the pH sensor is not submerged in storage solution, 
calibration solution, or water. Once the layer breaks down or is 
destroyed, the pH sensor will not provide accurate readings. 

When the pH sensor is properly kept in storage solution or 
deployed in water, the hydration layer can last for 6 months. 
The hydration layer then naturally breaks down and will result 
in unstable readings. Therefore, install a new pH electrode 
every 6 months. 

Important:

 New pH electrodes are uncalibrated. Once a pH 

electrode is installed in the logger for the first time, it must be 
calibrated before configuring and deploying the logger (see 

Configuring the Logger

). 

To install the pH electrode in the logger: 

1. Unscrew the closure cap over the sensor end cap of the

logger. Pull the sensor end cap straight out from the logger
without twisting it.

2. If this is the first time you’re installing a pH electrode, skip

to step 3. If a pH electrode is currently installed, gently
remove it from the sensor end cap. Use a foam swab
dipped in water to clean the internal O-rings. Use the foam
swab to spread a small dot of silicone-based grease along
both the O-rings inside the sensor end cap. Do not use a
cotton swab as the fibers on the swab can interfere with
the operation of the electrode.

3. Remove the pH electrode from the shipping bottle and wipe

the outer plastic body dry with a soft paper or cloth towel.

Never touch the glass bulb inside the clear plastic end of
the probe.

4. Remembering not to touch the glass bulb, gently push the

pH electrode into the sensor end cap. You will feel the pH
electrode pause as it passes through two internal O-rings
inside the sensor end cap (shown with step 2). Make sure
the internal O-rings are still in place as the probe is pushed
into the sensor end cap. Gently twist the electrode so that
the holes in the clear electrode body line up with the holes
in the end cap. Continue to gently push it into place until
the clear plastic is fully seated in the end cap as shown.

 

5. Push the sensor end cap back into the logger, making sure

the small tab in the end cap fits in the small groove in the
logger housing. Use the large tab on the end cap and the
arrow on the logger housing to help with proper alignment.
Do not twist or turn the end cap as you push it straight into
position.

6. Screw the closure cap back on the logger. 

7. Place the electrode into the storage solution. Slide the

storage solution cap with the rubber grommet on the
sensor end of the logger. Remove the cap from the storage
solution bottle and insert the sensor into the bottle. Screw

pH electrode 

pH sensor 
in storage 

solution 

First, unscrew this closure cap 

Then, pull this 
sensor end cap 
straight out 

Gently push pH electrode 
into end cap and twist the 
electrode so the holes in 
the electrode align with the 
holes in the end cap 

pH electrode installed 

Use a foam swab dipped in 
water to clean the internal 
O-rings; use a foam swab 
to spread a dot of silicone-
based grease on the 
internal O-rings 

Push the end cap 
back into logger, 
aligning tabs with 
arrow 

Sensor end cap 
installed with 
correct alignment 

Large tab 

Small tab

Small groove 

Summary of Contents for MX series

Page 1: ...ately 30 5 m 100 ft line of sight in air Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval normal statistics or burst Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory is full push button date time or after a set logging period Time Accuracy 1 min...

Page 2: ...cap can detect the presence of water This allows you to configure the logger in a power saving mode in which Bluetooth advertising is active only when the logger is removed from water See Configuring the Logger for details Note The logger will check for the presence of water every 15 seconds when the Bluetooth Off Water Detect power saving mode is selected pH Electrode This is the electrode that i...

Page 3: ...ut from the logger without twisting it 2 If this is the first time you re installing a pH electrode skip to step 3 If a pH electrode is currently installed gently remove it from the sensor end cap Use a foam swab dipped in water to clean the internal O rings Use the foam swab to spread a small dot of silicone based grease along both the O rings inside the sensor end cap Do not use a cotton swab as...

Page 4: ... Configure Select logger settings and load them onto the logger to start logging See Configuring the Logger Readout Download logger data See Reading Out the Logger pH Calibration Calibrate the logger following the instructions in HOBOmobile The logger should be calibrated after replacing or cleaning the electrode See Calibrating the Logger Full Status Details Check the battery level and view the c...

Page 5: ... as prompted making sure the sensor end cap temperature sensor and closure cap are submerged in the solution Follow the instructions on the screen Once the pH reading is stable tap Confirm Buffer 10 If you selected three point calibration repeat steps 8 9 11 Tap Save Calibration 12 Rinse the pH sensor with deionized or distilled water again If you will not be deploying the logger immediately attac...

Page 6: ...e Setting up Alarms for more details 7 Tap Start Logging and select when logging will begin Now Logging will begin immediately after tapping Start in the Configure screen On Next Logging Interval Logging will begin at the next even interval as determined by the selected logging interval On Button Push Logging will begin once you press the switch on the logger for 3 seconds On Date Time Logging wil...

Page 7: ...the High Alarm if you want an alarm to trip when the sensor reading rises above the high alarm value Drag the slider to the reading that will trip the alarm or tap the value field and type a specific reading In the example an alarm will trip when the temperature rises above 85 F 5 Enable the Low Alarm if you want an alarm to trip when the sensor reading falls below the low alarm value Drag the sli...

Page 8: ...ensor Limits 5 Enable High Limit if you want burst logging to occur when the sensor reading rises above a specific reading Drag the slider to the reading that will trigger burst logging or tap the value field and type a specific reading In this example the logger will switch to burst logging when the temperature rises above 85 F 6 Enable Low Limit if you want burst logging to occur when the sensor...

Page 9: ...n select Fixed Interval Logging 4 Select Normal to record the current reading for each enabled sensor at the logging interval shown at the top of the screen Do not select this if you only want to log statistics 5 Select the statistics you want the logger to record at each logging interval Maximum Minimum Average and Standard Deviation average is automatically enabled when selecting Standard Deviat...

Page 10: ...ed because the reading was outside the alarm limits or back within range Note Although the reading may have returned to a normal range during logging an alarm cleared event will not be logged if the logger was set up to maintain alarms until reconfigured New Interval The logger has switched to logging at the burst logging rate or back to the normal rate Water Detect The logger has been placed in w...

Page 11: ...s on the sensor end cap shown above Using your finger spread a small dot of grease along both outer O rings Battery Information The logger requires one user replaceable AA 1 5 V alkaline or lithium battery With a logging interval of 1 minute and at 25 C 77 F battery life is 1 year with Bluetooth Always On selected 2 years with Bluetooth Off Water Detect selected and 3 years with Bluetooth Always O...

Page 12: ...nto position 9 Screw the closure cap back on the logger WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case Do not dispose of the logger or battery in fire Do not expose the contents of the battery to water Dispose of the battery accordin...

Page 13: ...dustry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Avis de conformité pour l Industrie Canada Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts ...

Reviews: