background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

4

the storage solution cap installed on the logger onto the 
bottle as shown below. 

Important:

 Keep the sensor in this storage solution bottle 

unless you are calibrating the logger or deploying it in water. 
Only use storage solution in the bottle; do not use water. Do 
not use the copper guard while the sensor is in the bottle. 

Downloading HOBOmobile and  
Connecting to a Logger 

Install the HOBOmobile app to connect to and work with the 
logger. 

1. Download the HOBOmobile app from the App Store®.

2. Open the app and enable Bluetooth in the device settings if

prompted. 

3. Press the button on the logger to wake it up.

4. Tap the HOBOs icon at the bottom of the screen. Tap the

logger in the list to connect to it.

If the logger does not appear in the list or if it is having trouble 
connecting, follow these tips: 

If the logger was configured with Bluetooth Always Off
(see 

Configuring the Logge

r), make sure the logger is

“awake” by pressing the switch for 1 second. This will
also bring the logger to the top of the list if you are
working with multiple loggers.

Make sure the logger is within range of your mobile

device. The range for successful wireless communication
is approximately 30.5 m (100 ft) in air with full line-of-
sight.

Change the orientation of your phone or tablet to ensure

the antenna in your device is pointed toward the logger.
Obstacles between the antenna in the device and the
logger may result in intermittent connections.

If the logger is in water and configured with Bluetooth
Off Water Detect, you must remove the logger from the
water to connect to it.

If your device can connect to the logger intermittently or

loses its connection, move closer to the logger, within
sight if possible. If the logger is in water, the connection
can be unreliable. Remove it from water for a consistent
connection.

If the logger appears in the list, but you cannot connect
to it, close HOBOmobile and power cycle the mobile

device. This forces the previous Bluetooth connection to 
close. 

Once connected to the logger you can select one of the 
following actions: 

Configure.

 Select logger settings and load them onto the

logger to start logging. See 

Configuring the Logger

.

Readout.

 Download logger data. See 

Reading Out the

Logger

.

pH Calibration.

 Calibrate the logger following the

instructions in HOBOmobile. The logger should be
calibrated after replacing or cleaning the electrode. See

Calibrating the Logger

.

Full Status Details.

 Check the battery level and view the

configuration settings currently selected for the logger.

Start Logging. 

Select this option to begin logging (if the

logger is configured to start “On Button Push” as
described in 

Configuring the Logger

).

Stop Logging.

 Stop the logger from recording data. This

overrides any Stop Logging settings described in

Configuring the Logger

.

Page Logger LED.

 Press and hold this option to illuminate

the logger LEDs for 5 seconds.

Logger Password.

 Select this to create a password for the

logger that will be required if another mobile device
attempts to connect to it. To reset a password, connect
to the logger, tap Set Logger Passkey, and select Reset to
Factory Default. You can also press the button on the
logger for 10 seconds to reset a password. 

Update Firmware.

 When new logger firmware is

available, this action appears in the list. Select it and
follow the instructions on the screen. A logger readout
will be completed automatically at the beginning of the
firmware update process. If the connection is lost
between the logger and the mobile device during the
firmware update, a Firmware Update Pending Status
displays for the logger in the HOBOs list. Connect to
the logger and select Restore Logger (or Update
Firmware if that option is available) to continue
updating the firmware. 

Note:

 The logger will need to

be recalibrated after updating the firmware.

Important:

 Before updating the firmware on the 

logger, check the remaining battery level by selecting 
Full Status Details and make sure it is no less than 30%. 
Make sure you have the time to complete the entire 
update process, which requires that the logger remains 
connected to the device during the upgrade. 

Force Offload.

 This may appear if an error was

encountered when loading configuration settings. Select
this to offload all the data on the logger before
reconfiguring it.

First, slide on storage solution cap 

Then, insert sensor end cap in the bottle 

Summary of Contents for MX series

Page 1: ...ately 30 5 m 100 ft line of sight in air Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval normal statistics or burst Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory is full push button date time or after a set logging period Time Accuracy 1 min...

Page 2: ...cap can detect the presence of water This allows you to configure the logger in a power saving mode in which Bluetooth advertising is active only when the logger is removed from water See Configuring the Logger for details Note The logger will check for the presence of water every 15 seconds when the Bluetooth Off Water Detect power saving mode is selected pH Electrode This is the electrode that i...

Page 3: ...ut from the logger without twisting it 2 If this is the first time you re installing a pH electrode skip to step 3 If a pH electrode is currently installed gently remove it from the sensor end cap Use a foam swab dipped in water to clean the internal O rings Use the foam swab to spread a small dot of silicone based grease along both the O rings inside the sensor end cap Do not use a cotton swab as...

Page 4: ... Configure Select logger settings and load them onto the logger to start logging See Configuring the Logger Readout Download logger data See Reading Out the Logger pH Calibration Calibrate the logger following the instructions in HOBOmobile The logger should be calibrated after replacing or cleaning the electrode See Calibrating the Logger Full Status Details Check the battery level and view the c...

Page 5: ... as prompted making sure the sensor end cap temperature sensor and closure cap are submerged in the solution Follow the instructions on the screen Once the pH reading is stable tap Confirm Buffer 10 If you selected three point calibration repeat steps 8 9 11 Tap Save Calibration 12 Rinse the pH sensor with deionized or distilled water again If you will not be deploying the logger immediately attac...

Page 6: ...e Setting up Alarms for more details 7 Tap Start Logging and select when logging will begin Now Logging will begin immediately after tapping Start in the Configure screen On Next Logging Interval Logging will begin at the next even interval as determined by the selected logging interval On Button Push Logging will begin once you press the switch on the logger for 3 seconds On Date Time Logging wil...

Page 7: ...the High Alarm if you want an alarm to trip when the sensor reading rises above the high alarm value Drag the slider to the reading that will trip the alarm or tap the value field and type a specific reading In the example an alarm will trip when the temperature rises above 85 F 5 Enable the Low Alarm if you want an alarm to trip when the sensor reading falls below the low alarm value Drag the sli...

Page 8: ...ensor Limits 5 Enable High Limit if you want burst logging to occur when the sensor reading rises above a specific reading Drag the slider to the reading that will trigger burst logging or tap the value field and type a specific reading In this example the logger will switch to burst logging when the temperature rises above 85 F 6 Enable Low Limit if you want burst logging to occur when the sensor...

Page 9: ...n select Fixed Interval Logging 4 Select Normal to record the current reading for each enabled sensor at the logging interval shown at the top of the screen Do not select this if you only want to log statistics 5 Select the statistics you want the logger to record at each logging interval Maximum Minimum Average and Standard Deviation average is automatically enabled when selecting Standard Deviat...

Page 10: ...ed because the reading was outside the alarm limits or back within range Note Although the reading may have returned to a normal range during logging an alarm cleared event will not be logged if the logger was set up to maintain alarms until reconfigured New Interval The logger has switched to logging at the burst logging rate or back to the normal rate Water Detect The logger has been placed in w...

Page 11: ...s on the sensor end cap shown above Using your finger spread a small dot of grease along both outer O rings Battery Information The logger requires one user replaceable AA 1 5 V alkaline or lithium battery With a logging interval of 1 minute and at 25 C 77 F battery life is 1 year with Bluetooth Always On selected 2 years with Bluetooth Off Water Detect selected and 3 years with Bluetooth Always O...

Page 12: ...nto position 9 Screw the closure cap back on the logger WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case Do not dispose of the logger or battery in fire Do not expose the contents of the battery to water Dispose of the battery accordin...

Page 13: ...dustry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Avis de conformité pour l Industrie Canada Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts ...

Reviews: