background image

HOBO MX pH and Temperature Logger (MX2501) Manual 

7

choosing statistics options). In burst mode, logging occurs at 
a different interval when a specified condition is met. See 

Burst Logging

 for more information. Tap Done. 

11. Enable or disable Show LED. If Show LED is disabled, the

alarm and status LEDs on the logger will not be illuminated
while logging (the alarm LED will not blink if an alarm trips).
You can temporarily turn on LEDs when Show LED is
disabled by pressing the button on the logger for 1 second. 

12. Select the power saving mode, which determines when the

logger will “advertise” or regularly send out a Bluetooth
signal for the phone or table to find via HOBOmobile.

Bluetooth Always Off.

 The logger will only advertise

during logging when you push the switch on the logger.
This will wake up the logger when you need to connect to
it. This option uses the least battery power.

Bluetooth Off Water Detect. 

The logger will not

advertise when the presence of water is detected. Once
the logger is removed from the water, advertising will
automatically turn on, thereby not requiring you to push
the switch to wake up the logger when you need to
connect to it. This option preserves some battery power.

Note:

 The logger will check for the presence of water

every 15 seconds when this is option is selected.

Bluetooth Always On.

 The logger will always advertise.

You will never need to push a switch to wake up the
logger. This option uses the most battery power.

13. Tap Start in the upper right corner of the Configure screen

to load the settings onto the logger.

Logging will begin based on the settings you selected. See 

Deploying and Mounting the Logger

 for details on mounting 

and see 

Reading Out the Logger 

for details on downloading.

 

Setting up Alarms 

You can set up alarms for the logger so that if a sensor reading 
rises above or falls below a specified value, the red logger LED 
will blink and an alarm icon will appear in HOBOmobile. This 
can alert you to problems so you can take corrective action. 

To set an alarm: 

1. Tap the HOBOs icon and tap the logger to connect to it. If

the logger was configured with Bluetooth Always Off
enabled, push the switch on the logger to wake it up. If the
logger was configured with Bluetooth Off Water Detect and
is currently underwater, remove it from the water.

2. Once connected, tap Configure. 

3. In Sensor & Alarm Setup, tap Temperature or pH.

4. Enable the High Alarm if you want an alarm to trip when the

sensor reading rises above the high alarm value. Drag the
slider to the reading that will trip the alarm or tap the value
field and type a specific reading. In the example, an alarm
will trip when the temperature rises above 85°F.

5. Enable the Low Alarm if you want an alarm to trip when the

sensor reading falls below the low alarm value. Drag the
slider to the reading that will trip the alarm or tap the value
field and type a specific reading. In the example, an alarm is
configured to trip when the temperature falls below 32°F.

Note:

 The actual values for the high and low alarm limits are

set to the closest value supported by the logger. 

6. Under Raise Alarm After, select the duration before an

alarm is tripped and tap Done in the Alarm Duration screen. 

7. Select either Cumulative or Consecutive Samples. If you

select Cumulative Samples, then the alarm will trip when
the time the sensor is out of range over the course of the
deployment is equal to the selected duration. If you select
Consecutive Samples, then the alarm will trip when the
time the sensor is continuously out of range is equal to the
selected duration. For example, the high alarm for
temperature is set to 85°F and the duration is set to 30
minutes. If Cumulative is selected, then an alarm will trip
once a sensor reading has been at or above 85°F for a total
of 30 minutes since the logger was configured; specifically,
this could be 15 minutes above 85°F in the morning and
then 15 minutes above 85°F again in the afternoon. If
Consecutive is selected, then an alarm will trip only if all
sensor readings are 85°F or above for a continuous 30-
minute period.

8. Tap Done and repeat steps 3–8 for the other sensor if

desired. Note that when both alarms are configured, an
alarm is raised when either sensor is in an alarm condition.

9. Back in the Configure screen, select one of the following

options to determine how the alarm indications are cleared:

Logger Reconfigured.

 The alarm indication will display

until the next time the logger is reconfigured.

Sensor in Limits.

 The alarm icon indication will display

until the sensor reading returns to the normal range
between any configured high and low alarm limits.

10. Tap Start in the Configure screen to load the alarm settings

onto the logger if you are ready to start.

When an alarm trips, the red logger LED blinks every 4 seconds 
(unless Show LED is disabled), an alarm icon appears in the app, 
and an Alarm Tripped event is logged. The alarm state will clear 
when the readings return to normal if you selected Sensor in 

Summary of Contents for MX series

Page 1: ...ately 30 5 m 100 ft line of sight in air Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval normal statistics or burst Memory Modes Wrap when full or stop when full Start Modes Immediate push button date time or next interval Stop Modes When memory is full push button date time or after a set logging period Time Accuracy 1 min...

Page 2: ...cap can detect the presence of water This allows you to configure the logger in a power saving mode in which Bluetooth advertising is active only when the logger is removed from water See Configuring the Logger for details Note The logger will check for the presence of water every 15 seconds when the Bluetooth Off Water Detect power saving mode is selected pH Electrode This is the electrode that i...

Page 3: ...ut from the logger without twisting it 2 If this is the first time you re installing a pH electrode skip to step 3 If a pH electrode is currently installed gently remove it from the sensor end cap Use a foam swab dipped in water to clean the internal O rings Use the foam swab to spread a small dot of silicone based grease along both the O rings inside the sensor end cap Do not use a cotton swab as...

Page 4: ... Configure Select logger settings and load them onto the logger to start logging See Configuring the Logger Readout Download logger data See Reading Out the Logger pH Calibration Calibrate the logger following the instructions in HOBOmobile The logger should be calibrated after replacing or cleaning the electrode See Calibrating the Logger Full Status Details Check the battery level and view the c...

Page 5: ... as prompted making sure the sensor end cap temperature sensor and closure cap are submerged in the solution Follow the instructions on the screen Once the pH reading is stable tap Confirm Buffer 10 If you selected three point calibration repeat steps 8 9 11 Tap Save Calibration 12 Rinse the pH sensor with deionized or distilled water again If you will not be deploying the logger immediately attac...

Page 6: ...e Setting up Alarms for more details 7 Tap Start Logging and select when logging will begin Now Logging will begin immediately after tapping Start in the Configure screen On Next Logging Interval Logging will begin at the next even interval as determined by the selected logging interval On Button Push Logging will begin once you press the switch on the logger for 3 seconds On Date Time Logging wil...

Page 7: ...the High Alarm if you want an alarm to trip when the sensor reading rises above the high alarm value Drag the slider to the reading that will trip the alarm or tap the value field and type a specific reading In the example an alarm will trip when the temperature rises above 85 F 5 Enable the Low Alarm if you want an alarm to trip when the sensor reading falls below the low alarm value Drag the sli...

Page 8: ...ensor Limits 5 Enable High Limit if you want burst logging to occur when the sensor reading rises above a specific reading Drag the slider to the reading that will trigger burst logging or tap the value field and type a specific reading In this example the logger will switch to burst logging when the temperature rises above 85 F 6 Enable Low Limit if you want burst logging to occur when the sensor...

Page 9: ...n select Fixed Interval Logging 4 Select Normal to record the current reading for each enabled sensor at the logging interval shown at the top of the screen Do not select this if you only want to log statistics 5 Select the statistics you want the logger to record at each logging interval Maximum Minimum Average and Standard Deviation average is automatically enabled when selecting Standard Deviat...

Page 10: ...ed because the reading was outside the alarm limits or back within range Note Although the reading may have returned to a normal range during logging an alarm cleared event will not be logged if the logger was set up to maintain alarms until reconfigured New Interval The logger has switched to logging at the burst logging rate or back to the normal rate Water Detect The logger has been placed in w...

Page 11: ...s on the sensor end cap shown above Using your finger spread a small dot of grease along both outer O rings Battery Information The logger requires one user replaceable AA 1 5 V alkaline or lithium battery With a logging interval of 1 minute and at 25 C 77 F battery life is 1 year with Bluetooth Always On selected 2 years with Bluetooth Off Water Detect selected and 3 years with Bluetooth Always O...

Page 12: ...nto position 9 Screw the closure cap back on the logger WARNING Do not cut open incinerate heat above 85 C 185 F or recharge the lithium battery The battery may explode if the logger is exposed to extreme heat or conditions that could damage or destroy the battery case Do not dispose of the logger or battery in fire Do not expose the contents of the battery to water Dispose of the battery accordin...

Page 13: ...dustry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Avis de conformité pour l Industrie Canada Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts ...

Reviews: