background image

HOBO MX Water Level Logger (MX2001-0x) Manual

 

U.S. and International Sales: 1-508-759-9500 7 

www.onsetcomp.com 

General Guidelines for Deploying the Logger 

 

The absolute pressure sensor is temperature 
compensated over the range of 0° to 40°C (32° to 104°F). 
The barometric pressure sensor is temperature 
compensated over the range of -20° to 50°C (-4° to 
122°F). To obtain the highest level of accuracy, both the 
sensor end and top end units of the logger should be 
allowed to come to full temperature equilibrium 
(approximately 20 minutes) before you start the logger 
measure the reference level. 

 

Sudden temperature changes should be avoided. 

 

The sensor face located in the nose cone of the sensor 
end needs to be in the water to measure water level. 

 

 

Any change in length of the logger cable will result in a 1-
to-1 corresponding error in the depth measurement. 
Always pull-test a cable prior to deploying a logger in a 
well to make sure it does not stretch. 

 

To prevent the sensor from moving in currents and to 
ensure the support cable is kept straight during 
deployment, you may need to add a weight to the 
suspension cable just above the sensor or hang a weight 
below the sensor. In some cases, you may need to both 
add a weight and use a stilling well. 

 

Be very careful not to exceed the burst pressure for the 
sensor. The pressure sensor will burst if the maximum 
depth is exceeded (see 

Specifications

). The sensor should 

be positioned at a depth where it will remain in the water 
for the

 

duration of the deployment, but not exceed the 

rated bursting depth. 

 

If the environment where you are deploying the logger is 
known to have ants that are attracted to electronics, take 
appropriate protective measures to protect the logger, 
such as using pesticides. Even though the vent 
membrane for the logger’s barometric sensor is 
protected with a screen, some ants have eaten through 
this screen and the membrane, creating a path for 
moisture into the logger. 

 

Avoid deploying the top-end unit in direct sun as that can 
lead to more rapid temperature swings, which can lead 
to more measurement error and possibly hurt logger 
reliability. 

To deploy the logger: 

1.

 

Make sure the top end is connected to the sensor end with 
the water level logger cable as described in 

Getting Started 

with the Logger

2.

 

Attach any small parts, such as bolts with nuts, before you 
go into the field so you don’t risk losing them in the field. 

3.

 

Lower the logger into the well or stilling well.  

4.

 

Use the Onset well cap if it is a 5 cm (2 inch) well. 
Otherwise, use a 1/4-inch bolt of sufficient length to span 
the well top or washer. Mount the logger as follows: 

 

The top end must be above any metal well to ensure 
good wireless transmission. 

 

Make sure there is access for measuring the water level 
in the well. In some cases, such as when using the Onset 
well cap, it will be necessary to pull the top end of the 
logger out of the well to get a water level meter sensor 
into the well. 

 

Make sure the logger cannot accidentally fall in the well. 

 

If the cable is too long, loop the cable and secure the 
cable with multiple zip ties to ensure the loop does not 
slip. The looped cable should be tight enough that the 
cable can be easily pulled out of the well if necessary, but 
it must not bend the cable any tighter than a 1.25 cm (0.5 
inch) radius to prevent damage to the cable. 

5.

 

Measure the water level from the desired reference point 
(top of pipe, ground level, or sea level) and enter it in the 
app. Note that you may need to pull the top end out of the 
well to gain access for measuring the water level within the 
well. Follow these guidelines for determining the reference 
level: 

 

To maximize accuracy, allow 20 minutes after deploying 
the logger before starting the logger and measuring 
water depth to allow the logger to reach temperature 
equilibrium with the water. 

 

If the well is too small in diameter to measure the water 
depth while the logger is deployed, measure the water depth 
before deployment, then deploy the logger immediately. 

Make sure the 
sensor face 
(represented by 
dashed lines in 
this diagram) is in 
the water to 
measure water 
level 

10.0 mm (0.39 in.) 

Summary of Contents for MX2001-0 Series

Page 1: ... 0 02 kPa 0 003 psi 0 21 cm 0 007 ft water Pressure Response Time 90 1 second at a stable temperature Pressure Absolute and Water Level Measurements MX2001 02 SS S Operation Range 0 to 400 kPa 0 to 58 psia approximately 0 to 30 6 m 0 to 100 ft of water depth at sea level or 0 to 33 6 m 0 to 111 ft of water at 3 000 m 10 000 ft of altitude Factory Calibrated Range 69 to 400 kPa 10 to 58 psia 0 to 4...

Page 2: ... 1 C at 25 C 0 18 F at 77 F see Plot A Response Time 90 5 minutes in water typical Stability Drift 0 1 C 0 18 F per year Logger Operating Range 20 to 50 C 4 to 122 F Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 second to 18 hours Logging Modes Fixed interval multiple intervals with up ...

Page 3: ... to achieve full temperature compensation of the pressure sensor There can be up to 0 5 of additional error due to rapid temperature changes Measurement accuracy also depends on temperature response time Note Read all the directions in the Deploying a Logger section before deploying the logger Logger Components and Operation Disassembled top end The MX water level logger consists of a top end and ...

Page 4: ...layed as when the logger is stopped 6 Deploy the logger if ready If you are ready to deploy the logger place it in the deployment location see Deploying the Logger for deployment guidelines Otherwise skip deployment and complete the following steps as a test 7 Connect to the logger Tap the logger tile in the app to connect to the logger Tips for connecting Make sure the top end of the logger is wi...

Page 5: ...he current water level when the logger starts logging make sure the sensor is in position when logging starts Make this reading as accurate as possible since this reading is used to calibrate the water level data and adjust it to your reference See section 5 starting on page 7 for information on how to measure and enter the reference water level The logger must be logging for you to enter the refe...

Page 6: ...grommets are available at www onsetcomp com part number GROMMET 5 To install the grommet with the logger 1 Drill two 3 mm or 1 8 inch holes on opposite sides of the well cap for cross ventilation The holes should be just above the top of the grommet when installed as shown 2 Install the grommet with the upper ring sitting on top of the well as shown 3 Install the well cap on top of the well refer ...

Page 7: ...the top end unit in direct sun as that can lead to more rapid temperature swings which can lead to more measurement error and possibly hurt logger reliability To deploy the logger 1 Make sure the top end is connected to the sensor end with the water level logger cable as described in Getting Started with the Logger 2 Attach any small parts such as bolts with nuts before you go into the field so yo...

Page 8: ... alarm is set to 85 F and the duration is set to 30 minutes then the alarm will trip once the sensor readings have been above 85 F for a total of 30 minutes since the logger was configured Consecutive The alarm trips once the sensor reading is out of the acceptable range continuously for the selected duration For example the high alarm is set to 85 F and the duration is set to 30 minutes then the ...

Page 9: ... Logger Events for details on plotting and viewing the event Statistics Logging During fixed logging the logger records data for enabled sensors and or selected statistics at the selected logging interval Statistics are calculated at a sampling rate you specify with the results for the sampling period recorded at each logging interval You can log the following statistics for each sensor The maximu...

Page 10: ...wnloads data to the phone tablet or computer 3 Once the download is complete tap Done to return to the previous page or tap Export and Share to save the file in the specified format 4 When the export file has successfully been created tap Done to return to the previous page or tap Share to use your device s usual modes of sharing You can also upload data automatically to HOBOlink Onset s cloud sof...

Page 11: ...All pressure sensors drift over time The drift for the pressure sensors and electronics in the logger is less than 0 5 FS worst case per year In most applications drift is not a significant source of error because the offset created by any drift is zeroed out when you enter the reference water level in the app at the beginning of each deployment In effect you are re zeroing the sensor each time yo...

Page 12: ...rcuit board is normal when the batteries are installed Reinsert the board 6 Inspect the gore vent on the end cap to make sure it is not damaged Contact Onset or an Onset distributor if you need a replacement vent 7 Inspect the O rings and the mating surface on the logger for any damage clean off any debris and re lubricate the O rings with the grease You can leave the O rings on the end cap during...

Page 13: ...cations not expressly approved by the party responsible for compliance could void the user s authority to operate this equipment Industry Canada Statements This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may...

Reviews: