background image

 

HOBO Pendant MX Temp (MX2201) and Temp/Light (MX2202) Logger Manual 

1-800-LOGGERS 5 

field and type a specific reading. In the example, an alarm is 
configured to trip when the temperature falls below 32°F. 

6.

 

Under Raise Alarm After, select the duration before an 
alarm is tripped and tap Done in the Alarm Duration screen. 

7.

 

Select either Cumulative or Consecutive Samples. If you 
select Cumulative Samples, then the alarm will trip when 
the time the sensor is out of range over the course of the 
deployment is equal to the selected duration. If you select 
Consecutive Samples, then the alarm will trip when the 
time the sensor is continuously out of range is equal to the 
selected duration. For example, the high alarm for 
temperature is set to 85°F and the duration is set to 30 
minutes. If Cumulative is selected, then an alarm will trip 
once a sensor reading has been at or above 85°F for a total 
of 30 minutes since the logger was configured; specifically, 
this could be 15 minutes above 85°F in the morning and 
then 15 minutes above 85°F again in the afternoon. If 
Consecutive is selected, then an alarm will trip only if all 
sensor readings are 85°F or above for a continuous 30-
minute period. 

8.

 

Tap Done and repeat steps 3–8 for the other sensor in the 
Pendant MX Temp/Light (MX2202) logger if desired. Note 
that when both alarms are configured, an alarm is raised 
when either sensor is in an alarm condition. 

9.

 

Back in the Configure screen, select one of the following 
options to determine how the alarm indications are cleared: 

 

Logger Reconfigured.

 The alarm indication will display 

until the next time the logger is reconfigured. 

 

Sensor in Limits.

 The alarm icon indication will display 

until the sensor reading returns to the normal range 
between any configured high and low alarm limits. 

10.

 

Tap Start in the Configure screen to load the alarm settings 
onto the logger if you are ready to start. 

When an alarm trips, the logger alarm LED blinks every 4 
seconds (unless Show LED is disabled), an alarm icon appears in 
the app, and an Alarm Tripped event is logged. The alarm state 
will clear when the readings return to normal if you selected 
Sensor in Limits in step 9. Otherwise, the alarm state will 
remain in place until the logger is reconfigured. 

Notes: 

 

Alarm limits are checked at every logging interval. For 
example, if the logging interval is set to 5 minutes, then 
the logger will check the sensor readings against your 
configured high and low alarm setting every 5 minutes. 

 

The actual values for the high and low alarm limits are set 
to the closest value supported by the logger. For 
example, the closest value to 85°F that the logger can 
record is 84.990°F. In addition, alarms can trip or clear 
when the sensor reading is within the specified 
resolution. This means the value that triggers the alarm 
may differ slightly than the value entered. For example, if 
the High Alarm is set to 75.999°F, the alarm can trip 
when the sensor reading is 75.994°F, which is within the 
range of the resolution specifications. 

 

When you read out the logger, alarm events can be 
displayed on the plot or in the data file. See 

Logger 

Events

Burst Logging 

Burst logging is a logging mode that allows you to set up more 
frequent logging when a specified condition is met. For 
example, a logger is recording data at a 5-minute logging 
interval and burst logging is configured to log every 30 seconds 
when the temperature rises above 85°F (the high limit) or falls 
below 32°F (the low limit). This means the logger will record 
data every 5 minutes as long as the temperature remains 
between 85°F and 32°F. Once the temperature rises above 
85°F, the logger will switch to the faster logging rate and record 
data every 30 seconds until the temperature falls back to 85°F. 
At that time, logging then resumes every 5 minutes at the 
normal logging interval. Similarly, if the temperature falls below 
32°F, then the logger would switch to burst logging mode again 
and record data every 30 seconds. Once the temperature rises 
back to 32°F, the logger will then return to normal mode, 
logging every 5 minutes. 

Note:

 Sensor alarms, statistics, and the 

Stop Logging option “Wrap When Full” are not available in 
burst logging mode. 

To set up burst logging: 

1.

 

Tap the HOBOs icon and tap the logger to connect to it. If 
the logger was configured with Bluetooth Always Off 
enabled, press the circle on the logger to wake it up. 

2.

 

Once connected, tap Configure. 

3.

 

Tap Logging Mode and then tap Burst Logging. 

4.

 

Tap a sensor under Burst Sensor Limits. 

5.

 

Enable High Limit if you want burst logging to occur when 
the sensor reading rises above a specific reading. Drag the 
slider to the reading that will trigger burst logging or tap the 
value field and type a specific reading. In the example, the 
logger will switch to burst logging when the temperature 
rises above 85°F. 

 

6.

 

Enable Low Limit if you want burst logging to occur when 
the sensor reading falls below a specific reading. Drag the 
slider to the reading that will trigger burst logging or tap the 
value field and type a specific reading. In the example, the 
logger will switch to burst logging when the temperature 
falls below 32°F. 

7.

 

Tap Done and repeat steps 4–7 for the other sensor in the 
Pendant MX Temp/Light (MX2202) logger if desired. 

8.

 

Tap Burst Logging Interval and select an interval faster than 
the logging interval. Keep in mind that the faster the burst 
logging rate, the greater the impact on battery life and the 
shorter the logging duration. Because measurements are 
being taken at the burst logging interval throughout the 
deployment, the battery usage is similar to what it would be 
if you had selected this rate for the normal logging interval. 
Tap Done. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for Pendant MX Temp

Page 1: ...positive Waterproof To 30 5 m 100 ft Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 seco...

Page 2: ...stop logging Press this circle for 10 seconds to reset a password Note The circle on the front of the logger represents the button area on the logger You will not feel an actual button push when you p...

Page 3: ...ice attempts to connect to it To reset a password connect to the logger tap Set Logger Passkey and select Reset to Factory Default You can also press the circle on the logger for 10 seconds to reset t...

Page 4: ...interval selected see Statistics Logging for details on choosing statistics options In burst mode logging occurs at a different interval when a specified condition is met See Burst Logging for more i...

Page 5: ...hat triggers the alarm may differ slightly than the value entered For example if the High Alarm is set to 75 999 F the alarm can trip when the sensor reading is 75 994 F which is within the range of t...

Page 6: ...or lowest sampled value An average of all sampled values and The standard deviation from the average for all sampled values For example a Pendant MX Temp Light MX2202 logger is configured with both t...

Page 7: ...ng events to track logger operation and status To plot events in HOBOmobile tap a mini graph and then tap Select the events you wish to plot and then tap again You can also view events in shared or ex...

Page 8: ...nterval of 1 minute and Bluetooth Always Off disabled or 2 years typical at 25 C 77 F when the logger is configured with Bluetooth Always Off enabled Expected battery life varies based on the ambient...

Page 9: ...n If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interferen...

Reviews: