HOBO Pendant MX Temp (MX2201) and Temp/Light (MX2202) Logger Manual 

1-800-LOGGERS 6 

9.

 

Tap Done to exit the Logging Mode screen. 

10.

 

Tap Start in the Configure screen to load the burst settings 
onto the logger if you are ready to start. 

Notes: 

 

The high and low burst limits are checked at the burst 
logging interval rate whether the logger is in normal or 
burst condition. For example, if the logging interval is set 
to 1 hour and the burst logging interval is set to 10 
minutes, the logger will always check for burst limits 
every 10 minutes. 

 

If high and/or low limits have been configured for more 
than one sensor, then burst logging will begin when any 
high or low condition goes out of range. Burst logging will 
not end until all conditions on all sensors are back within 
normal range. 

 

The actual values for the burst logging limits are set to 
the closest value supported by the logger. For example, 
the closest value to 85°F that the logger can record is 
84.990°F. In addition, burst logging can begin or end 
when the sensor reading is within the specified 
resolution. This means the value that triggers burst 
logging may differ slightly than the value entered. For 
example, if the high limit for a temperature alarm is set 
to 75.999°F, burst logging can start when the sensor 
reading is 75.994°F, which is within the range of the 
resolution specifications. 

 

Once the high or low condition clears, the logging 
interval time will be calculated using the last recorded 
data point in burst logging mode, not the last data point 
recorded at the normal logging rate. For example, the 
logger has a 10-minute logging interval and logged a data 
point at 9:05. Then, the high limit was surpassed and 
burst logging began at 9:06. Burst logging then continued 
until 9:12 when the sensor reading fell back below the 
high limit. Now back in normal mode, the next logging 
interval will be 10 minutes from the last burst logging 
point, or 9:22 in this case. If burst logging had not 
occurred, the next data point would have been at 9:15. 

 

A New Interval event is created each time the logger 
enters or exits burst logging mode. See 

Logger Events

 for 

details on plotting and viewing the event. In addition, if 
the logger is stopped with a button push while in burst 
logging mode, then a New Interval event is automatically 
logged and the burst condition is cleared, even if the 
actual high or low condition has not cleared. 

Statistics Logging 

During fixed interval logging, the logger records data for 
enabled sensors and/or selected statistics at the logging 
interval selected. Statistics are calculated at a sampling rate you 
specify with the results for the sampling period recorded at 
each logging interval. The following statistics can be logged for 
each sensor: 

 

The maximum, or highest, sampled value, 

 

The minimum, or lowest, sampled value, 

 

An average of all sampled values, and 

 

The standard deviation from the average for all sampled 
values. 

For example, a Pendant MX Temp/Light (MX2202) logger is 
configured with both the temperature and light sensors 
enabled, and the logging interval set to 5 minutes. The logging 
mode is set to fixed interval logging with Normal and all four 
statistics enabled and with a statistics sampling interval of 30 
seconds. Once logging begins, the logger will measure and 
record the actual temperature and light values every 5 minutes. 
In addition, the logger will take a temperature and light sample 
every 30 seconds and temporarily store them in memory. The 
logger will then calculate the maximum, minimum, average, 
and standard deviation using the samples gathered over the 
previous 5-minute period and log the resulting values. When 
reading out the logger, this would result in 10 data series: two 
sensor series (with temperature and light data logged every 5 
minutes) plus eight maximum, minimum, average, and standard 
deviation series (four for temperature and four for light with 
values calculated and logged every 5 minutes based on the 30-
second sampling). 

To log statistics: 

1.

 

Tap the HOBOs icon and tap the logger to connect to it. If 
the logger was configured with Bluetooth Always Off 
enabled, press the circle on the logger to wake it up. 

2.

 

Once connected, tap Configure. 

3.

 

Tap Logging Mode and then select Fixed Interval Logging. 

4.

 

Select Normal to record the current reading for each 
enabled sensor at the logging interval shown at the top of 
the screen. Do not select this if you only want to log 
statistics. 

5.

 

Select the statistics you want the logger to record at each 
logging interval: Maximum, Minimum, Average, and 
Standard Deviation (average is automatically enabled when 
selecting Standard Deviation). Statistics will be logged for all 
enabled sensors. In addition, the more statistics you record, 
the shorter the logger duration and the more memory is 
required. 

6.

 

Tap Statistics Sampling Interval and select the rate to use 
for calculating statistics. The rate must be less than, and a 
factor of, the logging interval. For example, if the logging 
interval is 1 minute and you select 5 seconds for the 
sampling rate, then the logger will take 12 sample readings 
between each logging interval (one sample every 5 seconds 
for a minute) and use the 12 samples to record the resulting 
statistics at each 1-minute logging interval. Note that the 
faster the sampling rate, the greater the impact on battery 
life. Because measurements are being taken at the statistics 
sampling interval throughout the deployment, the battery 
usage is similar to what it would be if you had selected this 
rate for the normal logging interval. 

7.

 

Tap Done. 

8.

 

Tap Done again to exit the Logging Mode screen. 

9.

 

Tap Start in the Configure screen to load the statistics 
settings onto the logger if you are ready to start. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

sales@GlobalTestSupply.com

Summary of Contents for Pendant MX Temp

Page 1: ...positive Waterproof To 30 5 m 100 ft Radio Power 1 mW 0 dBm Transmission Range Approximately 30 5 m 100 ft line of sight Wireless Data Standard Bluetooth Low Energy Bluetooth Smart Logging Rate 1 seco...

Page 2: ...stop logging Press this circle for 10 seconds to reset a password Note The circle on the front of the logger represents the button area on the logger You will not feel an actual button push when you p...

Page 3: ...ice attempts to connect to it To reset a password connect to the logger tap Set Logger Passkey and select Reset to Factory Default You can also press the circle on the logger for 10 seconds to reset t...

Page 4: ...interval selected see Statistics Logging for details on choosing statistics options In burst mode logging occurs at a different interval when a specified condition is met See Burst Logging for more i...

Page 5: ...hat triggers the alarm may differ slightly than the value entered For example if the High Alarm is set to 75 999 F the alarm can trip when the sensor reading is 75 994 F which is within the range of t...

Page 6: ...or lowest sampled value An average of all sampled values and The standard deviation from the average for all sampled values For example a Pendant MX Temp Light MX2202 logger is configured with both t...

Page 7: ...ng events to track logger operation and status To plot events in HOBOmobile tap a mini graph and then tap Select the events you wish to plot and then tap again You can also view events in shared or ex...

Page 8: ...nterval of 1 minute and Bluetooth Always Off disabled or 2 years typical at 25 C 77 F when the logger is configured with Bluetooth Always Off enabled Expected battery life varies based on the ambient...

Page 9: ...n If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interferen...

Reviews: