56
Device functions
CALIBRATION
Calibration guarantees optimum performance of
the automatic function and mask test. It must be
performed when:
•
the device is used initially,
•
the mask has changed,
•
there is a change from operation without a
bacterial
fi
lter to with a bacterial
fi
lter, or vice
versa, or
•
there is a change from operation without a
humidi
fi
er to with a humidi
fi
er, or vice versa.
Calibration is only possible during mains opera-
tion and not in battery mode.
Calibrating the device
1. Connect the mask, the therapy tube and, if
applicable, the humidi
fi
er or bacterial
fi
lter to the
device.
Ensure that the air and exhalation openings in
the mask are clear and the tube is not kinked.
Do not put the mask on.