64
Device functions
BACT.FILTER
As pressure measurement is in
fl
uenced by the connec-
tion of a bacterial
fi
lter, it must be speci
fi
ed on the device
whether or not a bacterial
fi
lter has been connected.
Select whether a bacterial
fi
lter is connected (by
the physician)
1. Select the complete menu (see page 39).
2. Use the selection keys
and
to select
Bact
.
Filter
.
3. Press the the enter key
and if necessary,
input the PIN code with the selection keys
and
. Press the enter key
after every digit.
4. Press the enter key
and then use the selec-
tion keys
and
to set
Yes
or
No
. Con-
fi
rm your setting by pressing the enter key
.
Whenever a bacterial
fi
lter is connected to or
removed from the device, it is necessary to cali-
brate the device (see page 56).