34
Chapter 4: Commissioning
General information
•
Before commissioning the device, read the safety information from page 19.
•
Before commissioning the ventilation system (ventilator, tube, humidifier etc.), check
all connections for leaks, as well as the stability of the connected accessories.
•
Never operate the device without the air filter.
•
Only use original HOFFRICHTER filters.
•
If the device was previously in an environment where the air temperature was not
the same as in the new operating location, allow approximately 1 hour until the
temperatures have evened out before commissioning.
Setting up the device
Place the device on a flat and stable surface. Make sure that the device is placed securely
and that the air inlet at the rear of the device is not blocked. Make sure that the user
is able to see the display and the info LED during ventilation. The device is designed
for operation within arm’s reach.
FiO
2
FiO
2
> 10 cm
Figure 7:
Setting up the device
Summary of Contents for LAVI
Page 2: ......
Page 18: ...18 Chapter 1 Introduction ...
Page 26: ...26 Chapter 2 Safety warnings ...
Page 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Page 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Page 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Page 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Page 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Page 84: ...84 Chapter 8 Routine checks and maintenance work ...
Page 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Page 98: ...98 Chapter 9 Appendix ...
Page 99: ...Chapter 9 Appendix 99 ...
Page 100: ...100 Chapter 9 Appendix ...
Page 101: ...Chapter 9 Appendix 101 ...
Page 102: ...102 Chapter 9 Appendix ...