Chapter 4: Commissioning
If the calibration was not successful, “Error”
will appear.
In the event of an error, check the entire sys-
tem. Resistance in the overall system may
be too high. You may, for example, have to
exchange the bacterial filter or use another
humidifier. Then rerun the calibration.
PCV
Set 1
75%
50
40
30
20
10
0
p
- - . -
hPa
f
- -
bpm
V
I
- . - - -
l
I:E
- : - . -
Calibrate Tube
100 l/min
Error
Press the MFK to Confirm
If you do not perform a tube calibration, the calibration data from the last tube cali-
bration are used. When commissioning the device for the first time, the default cali-
bration data
1
stored in the device are used.
1
Operation without humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected ventilation components and accessories: no
Operation with humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected ventilation components and accessories: Humidifier AquaTREND uni
Summary of Contents for LAVI
Page 2: ......
Page 18: ...18 Chapter 1 Introduction ...
Page 26: ...26 Chapter 2 Safety warnings ...
Page 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Page 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Page 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Page 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Page 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Page 84: ...84 Chapter 8 Routine checks and maintenance work ...
Page 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Page 98: ...98 Chapter 9 Appendix ...
Page 99: ...Chapter 9 Appendix 99 ...
Page 100: ...100 Chapter 9 Appendix ...
Page 101: ...Chapter 9 Appendix 101 ...
Page 102: ...102 Chapter 9 Appendix ...