Chapter 6: Alarms and messages
Alarm test
The user does not have to perform any alarm tests. However, the following describes
an option to call the “Disconnection” alarm in order to perform a manual check of
the alarm system.
To check the “Disconnection” alarm:
1. Connect the device to the mains supply.
2. Switch the device on. Do not connect the leakage tube system!
3. Start ventilation.
4. After a few seconds, the “Disconnection”
alarm occurs. If this is not the case, the device
must be returned for servicing.
PCV
Set 1
75%
50
40
30
20
10
0
p
19.8
hPa
f
12
bpm
V
I
0.570
l
I:E
1:1.5
Statistics
Parameter
Event Log
Service
Monitoring
Thursday
I
19.05.2016
I
23:10
System
Disconnection
Summary of Contents for LAVI
Page 2: ......
Page 18: ...18 Chapter 1 Introduction ...
Page 26: ...26 Chapter 2 Safety warnings ...
Page 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Page 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Page 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Page 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Page 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Page 84: ...84 Chapter 8 Routine checks and maintenance work ...
Page 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Page 98: ...98 Chapter 9 Appendix ...
Page 99: ...Chapter 9 Appendix 99 ...
Page 100: ...100 Chapter 9 Appendix ...
Page 101: ...Chapter 9 Appendix 101 ...
Page 102: ...102 Chapter 9 Appendix ...