Chapter 9: Appendix
95
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity tests
IEC 60601 test level
Compliance level
Electromagnetic envi-
ronment - guidance
Voltage dips, tem-
porary interruptions
and voltage fluctua-
tions in the voltage
supply lines as per
IEC 61000-4-11
> 95 % dip of
U
T
for
1/2 period
> 95 % dip of
U
T
for 1
period
> 30 % dip of
U
T
for
25 (50 Hz) periods/
30 (60 Hz) periods
> 95 % dip of
U
T
for
250 (50 Hz) periods /
300 (60 Hz) periods
> 95 % dip of
U
T
for
1/2 period
> 60 % dip of
U
T
for
5 periods
> 30 % dip of
U
T
for
25 periods
> 95 % dip of
U
T
for
5s or 6s
The quality of the sup-
ply voltage should cor-
respond to one of the
typical business or hos-
pital environments. If the
user of the
LAVI
venti-
lator requires continued
functioning even in the
event of an interruption
to the power supply, we
recommended supplying
the
LAVI
ventilator using
an uninterruptable power
supply (UPS) or a battery.
Summary of Contents for LAVI
Page 2: ......
Page 18: ...18 Chapter 1 Introduction ...
Page 26: ...26 Chapter 2 Safety warnings ...
Page 27: ...Chapter 3 Description of device 27 Chapter 3 Description of Device ...
Page 33: ...Chapter 3 Description of device 33 Chapter 4 Commissioning ...
Page 51: ...Chapter 5 Device operation 51 Chapter 5 Operating the Device ...
Page 75: ...Chapter 6 Alarms and messages 75 Chapter 7 Cleaning and Exchange of Components ...
Page 81: ...Chapter 8 Routine checks and maintenance work 81 Chapter 8 Routine Tests and Maintenance Work ...
Page 84: ...84 Chapter 8 Routine checks and maintenance work ...
Page 85: ...Chapter 9 Appendix 85 Chapter 9 Annex ...
Page 98: ...98 Chapter 9 Appendix ...
Page 99: ...Chapter 9 Appendix 99 ...
Page 100: ...100 Chapter 9 Appendix ...
Page 101: ...Chapter 9 Appendix 101 ...
Page 102: ...102 Chapter 9 Appendix ...