Additional work
Action required
Page
Before every recommissioning
from 3 months of storage, checking the battery capacity with a
battery test
Safety-related check
Documentation
29
6
30
When changing patients, prepa-
ration, the preparation of con-
taminated devices (e.g., MRSA)
Hygienic preparation, in accordance with the TRENDvent hygiene
concept
Table 2: Additional work
3.2 REPAIRS
Repairs to the device may only be carried out by an authorized service agency.
The following should be carried out in the case of a repair:
•
Wipe disinfect the device surface, e.g., Mikrozid® Liquid
•
Check the error description and error analysis
•
Carry out the repairs
•
Final inspection consisting of:
– Safety-related test (see section 3.4 on page 6)
– Checking the electrical safety of the power supply (see section 8.6 on page 30)
– Final work (section 9 on page 30)
•
Packaging of the device according to the packing instructions in section 12.2.1 on page 35
•
Shipping according to the shipping instructions in section 12.2.2 on page 35
3.3 HYGIENIC PREPARATION
During preparation, preparation on changing patients, as well as preparation after contamination
(e.g., MRSA), the device should be treated according to the TRENDvent hygiene concept.
3.4 SAFETY-RELATED CHECK
The safety-related check must be carried annually, after repairs and before every recommissioning.
This consists of:
•
a visual inspection of the device (for external damage)
•
a functional check (see page 25)
•
a visual inspection of the accessories, in accordance with the manufacturer’s instructions (respi-
ratory tubes, oxygen sensor, etc.)
The safety-related check carried out must be documented.