Manufacturer’s declaration on electromagnetic compatibility
99
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic environment –
guidance
Voltage dips, short interruptions and
voltage variations on power supply
input lines acc. to IEC 61000-4-11
< 5 %
U
T
(> 95 %
dip in
U
T
) for 0.5
cycle
40 %
U
T
(60 % dip
in
U
T
) for 5 cycles
70 %
U
T
(30 % dip
in
U
T
) for 25 cycles
< 5 %
U
T
(> 95 %
dip in
U
T
) for 5 s
> 95 % dip in
U
T
for 0.5 cycle
60 % dip in
U
T
for
5 cycles
30 % dip in
U
T
for
25 cycles
> 95 % dip in
U
T
for 5 s
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the TRENDvent ventilator requires continued
operation during power mains interruption, it
is recommended that the TRENDvent ventila-
tor is powered from an uninterrupted power
supply (UPS) or a battery.
Power frequency (50/60 Hz) magnetic
fi
eld acc. to IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fi
elds should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
Conducted RF acc. to IEC 61000-4-6 3 V
effective value
150 kHz – 80 MHz
3 V
Portable and mobile communications equip-
ment should be used no closer to any part of
the TRENDvent ventilator, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended safety distance:
d = 1.16
P