Calibration instructions
geodyna 6300
Page 4 of 8
Date 2 Oct 2001
3 D SAPE Calibration of width gauge arm with code C82
Service and test functions become accessible by
simultaneously pressing and holding the C and the balancing
mode keys and by rotating the main shaft.
If the width gauge arm is in right home position and if the slider voltage of the
width potentiometer is not within a range of 4.25 to 4.30 Volt, turn the
potentiometer shaft to bring the voltage to within this range. Press the C key to
store the value.
The following reading appears:
Apply the gauge head of the width gauge to the outer surface of the adaptor
flange and press the C key to store the value.
The following reading appears:
Insert the calibration weight supplied with the machine into the adaptor flange
so that it protrudes to the right.
Apply the width gauge arm to the outer surface of the calibration weight and
press the C key to store the value.
The following reading appears:
Calibration is completed and must be stored by entering C90.
C82
1 4.26
2 0,77
3 2,33
C - -