Calibration instructions
geodyna 6300
Page 7 of 8
Date 2 Oct 2001
Calibration of residual unbalance of main shaft
and drive pulley using code C84
Service and test functions become accessible by simultaneously pressing and
holding the C and the balancing mode keys and by rotating the main shaft.
The following reading appears for one second:
Then the following reading appears:
Press the START key to start a measuring run.
Then the following reading appears:
Calibration is completed and must be stored by entering C90.
C84.
C84. Start
C - -