6-8 Duty cycle 

0.1%

99.9%: ± ( 2.0% of rdg + 2 digits ), Frequency lower than 10kHz 

-- Sensitivity: sine wave 0.6V rms 
-- Overload protection: 500V DC or AC rms 

6-9 Temperature

 

Range 

Accuracy 

Resolution 

 

-20~150

 

± ( 3

 

+ 1digit ) 

1

 

150~1000

 

± ( 3% of rdg + 2digits ) 

 

-4~302

 

± ( 5

 

+ 2digits ) 

1

 

302~1832

 

± ( 3% of rdg + 3digits ) 

-- NiCr-NiSi K-type sensor 
-- Overload protection: 800mA/500V Fast Fuse

 

6-10 Diode and Audible continuity test 

Range 

Description 

Test Condition 

Display read approximately 

forward voltage of diode 

Forward DC current 

approx. 1.5mA 

Reversed DC voltage 

approx. 3.2V 

Built-in buzzer sounds if 

resistance is less than 50Ω 

Open circuit voltage 

approx. 1V 

Overload protection: 500V DC or AC rms

 

6-11 Transistor hFE test 

Test range: 0-1000 

Ib=10µA, Vce=1.8V Approx. 

6-12 Non Contact AC Voltage (NCV) detection 

Test voltage range: 90V

1000V AC rms 

The NCV red light and green light will light up alternately together with sound. 

7. OPERATING INSTRUCTIONS

7-1 Attention before operation 

7-1-1 Check battery. When the battery voltage drop below proper operation range, the “

” symbol 

will appear on the LCD display and the battery need to changed. 
7-1-2 Pay attention to the “

” besides the input jack which shows that the input voltage or current 

should be within the specified value. 
7-1-3 The range switch should be positioned to desired range for measurement before operation. 

7-2 Measuring DC & AC Voltage 

7-2-1  Set  the  rotary  switch  at  the  desired “

mV

”  or “

V

”  range  position,  it  shows  symbol  for 

testing DC voltage, if you want to test AC voltage, push “SELECT” button switch. 
7-2-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-2-3 Connect test leads across the source or load under measurement. 
7-2-4 You can get reading from LCD. The polarity of the red lead connection will be indicated along 
with the DC voltage value. 

NOTE:

   

1.“

” means you can’t input the voltage more than 1000V DC or 750V AC, it’s possible to show 

higher voltage, but it may destroy the inner circuit or pose a shock. 
2. Be cautious against shock when measuring high Voltage.

7-3 Measuring DC & AC Current 

7-3-1  Set  the  rotary  switch  at  the  desired  “

uA

”  &  “

mA

”  &  “

A

”  range  position,  it  shows 

symbol for testing DC current, if you want to test AC current, push “SELECT” button switch. 
7-3-2 Connect the black test lead to 

COMT-

 jack and the red to the 

μAmAT+

 jack for a maximum 

600mA current , for a maximum 6A or 20A current, move the red lead to the 

20A 

jack. 

7-3-3 Connect test leads in series with the load under measurement. 
7-3-4 You can get reading from LCD. The polarity of the red lead connection will be indicated along 
with the DC current value. 

NOTE: 

1. When  the  value  scale  to  be  measured  is  unknown  beforehand,  set  the  range  selector  at  the 
highest position. 
2. When  only  “OL”  is  displayed,  it  indicates  over-range  situation  and  the  higher  range  has  to  be 
selected. 
3. “

” means the socket mA’s maximum current is 800mA and 20A’s maximum current is 20A, over 

800mA or 20A current can be protected by the fast fuse. 
4. On the 20A range, the measuring time should be less than 10 seconds to prevent precision from 
affecting by circuit heating. 

7-4 Measuring DC & AC 600A Current 

7-4-1 Set the rotary switch at the desired “

600A

” range position, it shows symbol for testing DC 

current, if you want to test AC current, push “SELECT” button switch. 
7-4-2 Connect the black banana plug of the AC/DC Current Clamp Adapter to 

COMT-

 jack and the 

red banana plug to the 

μAmAT+

 jack. 

7-4-3 Set the AC/DC Current Clamp Adapter to “

1mV/A

” range. 

7-4-4 When perform DC current measurement, always rotate or push the DCA zero adjuster on the 
Clamp Adapter until the multimeter reads zero. 
7-4-5 Clamp the Jaws around the 

one 

conductor to be measured. Center the conductor within the 

Jaw using the Centering Marks as guides. 
7-4-6 Read the result from the LCD panel. The arrow in the Jaw indicates the DC current direction of 
positive current flow (positive to negative). 

7-5 Measuring Resistance 

7-5-1 Set the rotary switch at the desired “

” range position. 

7-5-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-5-3 Connect test leads across the resistance under measurement. 
7-5-4 You can get reading from LCD. 

NOTE: 

Max. input overload: 500V rms

10sec

 

1. For measuring resistance above 1MΩ, the mete may take a few seconds to get stable reading.
2. When  the  input  is  not  connected,  i.e.  at  open  circuit,  the  figure  ‘OL’  will  be  displayed  for  the 
over-range condition. 
3. When checking in-circuit resistance, be sure the circuit under test has all power removed and that 
all capacitors have been discharged fully. 

7-6 Measuring Capacitance 

7-6-1 Set the rotary switch at the desired “

” range position, push “SELECT” to choose 

capacitance

 

measurement. 

7-6-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-6-3 Connect test leads across the capacitance under measurement. 
7-6-4 You can get reading from LCD. 

NOTE: 

Max. input overload: 500V rms

10sec

 

1.

Capacitors should be discharged before being tested.

2.

When testing large capacitance, it will take longer time before the final indication ( For 
100uF~99.99mF range, it will take about 10 seconds ). 

3.

When testing small capacitance (≤1uF), to assure the measurement accuracy, first press
"REL

", then go on measuring. 

7-7 Measuring Frequency & Duty cycle 

7-7-1 Set the rotary switch at the desired “

Hz

” range position. 

7-7-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-7-3 Push “

Hz/Duty

” key to choose 

Frequency 

or 

Duty cycle 

test. 

7-7-4 Connect the probe across the source or load under measurement. 
7-7-5 You can get reading from LCD. 

7-8 Measuring Temperature 

7-8-1 Set the rotary switch at the desired “

” range position, push “SELECT” to choose 

 

or

 

 

measurement. 
7-8-2 Connect the black banana plug of the sensor to 

COMT-

 jack and the red banana plug to the 

μAmAT+

 jack. 

7-8-3 Put the sensor probe into the temperature field under measurement. 
7-8-4 You can get reading from LCD. 

NOTE:

 

1.  The  accessory  of  the  meter  WRNM-010  type  contact  thermocouple  limit  temperature  is  250 

 

(300 

 shortly), please use special probe for test higher temperature. 

2. Please don't change the thermocouple at will, otherwise we can't guarantee to measure accuracy.
3. Please don’t importing the voltage in the temperature function.

7-9 Diode & Audible continuity Testing 

7-9-1 Set the rotary switch at the “

 

” range position, push “SELECT” to choose 

Diode 

or 

Audible continuity 

measurement. 

7-9-2 Connect the black test lead to 

COMT-

 jack and the red to 

VΩHz

 jack. 

7-9-3 On 

diode

 range, connect the test leads across the diode under measurement, display 

shows the approx. forward voltage of this diode. 
7-9-4 On 

Audible continuity

 range, connect the test leads to two point of circuit, if the resistance is 

lower than approx. 50Ω, the buzzer sounds. 

NOTE: 

Make  sure  the  power  is  cut  off  and  all  capacitors  need  to  be  discharged  under 

this measurement. 

7-10 Transistor hFE Test 

7-10-1 Set the rotary switch at the desired “

hFE

” range position. 

7-10-2 Determine whether the transistor is NPN or PNP and locate the Emitter, Base and 
Collector leads. Insert the leads into the proper holes in the socket on the front panel. 
7-10-3 You can get reading of the approximate hFE value from LCD. 

NOTE: 

Don’t connect an external voltage to measuring terminals. 

7-11 Non Contact AC Voltage detection 

7-11-1 Set the rotary switch at the desired “

NCV

” range position, the NCV green LED light light

 u

p. 

7-11-2 Hold the Meter so that the mater’s top is vertically and horizontally centered and contacting 
the conductor, when the live voltage ≥ 90V AC rms, the NCV red LED light and green LED light 
will light up alternately together with sound. 

NOTE: 

1.  Even  without  LED  indication,  the  voltage  may  still  exist.  Do  not  rely  on  non-contact  voltage 

detector to determine the presence of voltage wire. Detection operation may be subject to socket 

design, insulation thickness and different type and other factors. 

2.  When  the  meter  input  terminals  presence  voltage,  due  to  the  influence  of  presence  voltage, 

voltage sensing indicator may also be bright. 

3.  Keep  the  meter  away  from  electrical  noise  sources  during  the  tests,  i.e.,  florescent  lights, 

dimmable lights, motors, etc.. These sources can trigger Non-Contact AC Voltage detection function 

and invalidate the test. 

8. Battery replacement

8-1 When the battery voltage drop below proper operation range the "

" symbol will appear on the 

LCD display and the battery need to changed. 
8-2 Before changing the battery, set the selector switch to “

OFF

” position and remove the test leads 

from the terminals. Open the cover of the battery cabinet by a screwdriver. 
8-3 Replace the old battery with the same type battery (AA R6P 1.5V×2). 
8-4 Close the cover of the battery cabinet and fasten the screw. 

9. Fuse replacement

9-1  This  meter  is  provided  with  a  800mA/500V  fast  fuse  to  protect  the  temperature  test  and  the 

current measuring circuits which measure up to  600mA, with a 20A/500V fuse to protect the  20A 

range. 

9-2  Ensure  the  meter  is  not  connected  to  any  external  circuit,  set  the  selector  switch  to  “

OFF

” 

position and remove the test leads from the terminals. Open the cover of the battery cabinet  by a 

screwdriver. 

9-3 Replace the old fuse with the same type and rating: 6×30mm 800mA/500V fast fuse or 6×30mm 

20A/500V fast fuse. 

9-4 Close the cover of the battery cabinet and fasten the screw. 

10. Maintenance

10-1 You must replace the test leads if the  lead is exposed, and should adopt the leads with the 

same specifications as origin. 

10-2 Use only moist fabric or small amount of detergent but not chemical solution for cleaning. 

10-3 Do not use the meter before the back cover is properly closed and screw secured.

 

Upon any 

abnormality, stop operation immediately and send the meter for maintenance. 

10-4 Please take out the battery when not using for a long time. 