7
T
ROUBLESHOOTING
7.18
ThinPrep
®
3000 Processor Operator’s Manual
Out of Consumables
Reprocessing Required
Required consumable item(s) are not in place to process sample vials, consequently the sample vial
cannot be processed.
If the instrument detects that there are not enough consumable items to complete processing a sam-
ple, the “Items Must Be Addressed” message screen appears, listing the consumables required for
further processing.
You may follow the screen prompts and load supplies into the instrument or you may press
Cancel
to end the batch. The batch terminates and the last sample in process will not have a slide made. It is
listed on the batch report as Out of Consumables.
Possible Cause
• Instrument is out of ThinPrep Pap test filters.
• Instrument is out of slides, or cannot detect slide cartridges.
• Staining racks are full.
Corrective Action
1. Check and load items as prompted by the message screen.
Preventative Action
Items appear on the message screen after
Start Batch
is selected. Follow the message screen prompts
to load necessary items.
Summary of Contents for ThinPrep 3000
Page 1: ......
Page 4: ...The ThinPrep Processor The ThinPrep Processor ...
Page 5: ...MAN 03939 001 Rev 004 page 1 of 13 Instructions for Use ...
Page 18: ...Table of Contents Table of Contents ...
Page 23: ...1 Introduction 1 Introduction ...
Page 42: ...2 ThinPrep 3000 Installation 2 ThinPrep 3000 Installation ...
Page 55: ...3 PreservCyt and CellFyx Solutions 3 PreservCyt and CellFyx Solutions ...
Page 71: ...4 Sample Collection and Preparation 4 Sample Collection and Preparation ...
Page 80: ...5 Instrument Operation 5 Instrument Operation ...
Page 105: ...6 Maintenance 6 Maintenance ...
Page 148: ...7 Troubleshooting 7 Troubleshooting ...
Page 205: ...8 Staining and Coverslipping 8 Staining and Coverslipping ...
Page 212: ...9 The ThinPrep Pap Test Training Program 9 The ThinPrep Pap Test Training Program ...
Page 215: ...10 User Interface Screens 10 User Interface Screens ...
Page 226: ...Index Index ...
Page 232: ...INDEX Index 6 ThinPrep 3000 Processor Operator s Manual This page intentionally left blank ...
Page 233: ...Service Information Service Information ...
Page 236: ...Ordering Information Ordering Information ...
Page 243: ...Safety Data Sheets Safety Data Sheets ...
Page 246: ...Appendix Appendix ...
Page 255: ......
Page 256: ......