7.2 ETHERNET COMMUNICATION
7.2.1 LAN Connection
Communication with the HS9000B over a LAN connection defaults to the use of DHCP. The
instrument can be addressed by using either the network assigned IP address or by using the
instrument serial (ex. "HS9000B-123") and the
TCP port (9760)
. Use the Holzworth Ethernet Finder
software to locate and modify IP address settings on the instrument.
To search for devices, click the
Devices
button and then click
Locate Devices
in the sub menu.
The software will then scan for instruments connected via Ethernet and via serial port. It will display
Ethernet devices as shown below:
Identify the instrument by serial # or IP address and click to connect. If the connection is successful
the window above 'Devices' will turn green (Ethernet) and display the instrument serial number:
Users can also enter the instruments IP address manually to connect. Enter the IP address into the
'Device IP Address' field and then press the
Connect
button.
If the connection is successful the window above 'Devices' will turn green and display the IP
address.
7.2.2 DHCP
When the HS9000B is connected to a network with a DHCP server, the network settings will be
auto-assigned per the servers configuration.
If the HS9000B is connected
directly
to a PC or to a network with no DHCP server the instrument's
default IP address will be:
169.254.117.11