background image

©

 Home Training Tools Ltd. 2006                                        Page 5 of 8

        

                              Visit us at ww.homesciencetools.com 

 

Troubleshooting 

If you are experiencing difficulty with your microscope, try these troubleshooting techniques:  

Problem 

Possible Reason and Solution  

Light fails to 
operate 

1.

 

The batteries are dead (MI-1000LED). 

Use the AC adapter to recharge the batteries.

 

2.

 

The light intensity control is off (MI-1000LED). 

Turn up the light intensity. 

3.

 

The bulb is burned out. 

Replace the bulb. (See “Changing the Bulb,” p. 4.)

 

4.

 

The incorrect bulb is installed. 

Replace with the correct bulb. 

 

Light flickers 

1.

 

The bulb is not properly inserted into the socket. 

Properly insert the bulb.

 

2.

 

The bulb is about to burn out. 

Replace the bulb. 

 

No image 

1.

 

The nosepiece is not indexed properly. 

Move revolving nosepiece until the objective 

lens clicks into position.  

2.

 

The light is too bright. 

Adjust the diaphragm. 

Unable to 
focus slide 

1.

 

The slide coverslip is too thick. 

Use 0.17 mm thick (No. 1) coverslip. 

  

2.

 

The slide is upside down. 

Place the slide on the stage with the coverslip facing up.

 

3.

 

The stage stop is not set at the proper position. 

Adjust the stage stop. (See 

“Adjusting the Stage Stop,” p. 4.)

 

Poor 
resolution,  
image not 
sharp 

1.

 

The objective or eyepiece lenses are dirty. 

Clean the lenses. (See “Cleaning,” p. 2.)

 

2.

 

There is too much light. 

Adjust the diaphragm.

  

Spots in field 

1.

 

The specimen slide, objective, or eyepiece lens is dirty. 

Clean the slide or lenses. 

(See “Cleaning,” p. 2.)

 

Uneven 
illumination of 
field 

1.

 

The nosepiece is not indexed properly. 

Move revolving nosepiece until the objective 

lens clicks into position.  

2.

 

The diaphragm is not properly indexed. 

Adjust the diaphragm to the proper level. 

Specifications 

Eyepiece 

Widefield 10x eyepiece with fully coated optics. Inclined 45° head. 

Nosepiece 

3-hole, ball-bearing mounted with positive click stops. 

Objectives 

All objectives are achromatic, parfocalled, parcentered, and fully coated.  

 

4x, 0.10 N.A., red ring, 4.0mm field of view, 40x magnification 

 

10x, 0.25 N.A., yellow ring, 1.6mm field of view, 100x magnification 

 

40x, 0.65 N.A., blue ring, .40mm field of view, 400x magnification, retractable 

Focusing 

Single intermediate focusing control with slip clutch. All metal rack-and-pinion focusing with 
adjustable stage stop. 

Stage 

Acid and chemical resistant 95 x 95mm metal stage with stage clips (not designed for use 
with a mechanical stage).   

Condenser 

Fixed 0.65 NA condenser.  

Diaphragm 

Calibrated 6-hole disc diaphragm.  

Illuminator 

15-watt tungsten illuminator with grounded 110-volt cord on model MI-1000STD, 20-watt 
equivalent LED illuminator with AC Adapter or optional batteries on model MI-1000LED.  

Summary of Contents for MI-1000STD

Page 1: ...ruction Manual for MI 1000STD Kids Microscope MI 1000LED Kids LED Cordless Microscope 1 800 860 6272 www homesciencetools com Copyright 2006 by Home Training Tools Ltd All rights reserved Kids Microsc...

Page 2: ...up by the arm Avoid touching the lens surfaces on the eyepiece or objective lens as finger prints will decrease image quality Cleaning The best optical quality can be compromised by dirty lenses Using...

Page 3: ...size for example a hole labeled 6 is 6mm in diameter and a hole labeled 2 has a diameter of 2mm Use the smaller holes for lower magnification and the larger holes for higher magnification 9 Focus kno...

Page 4: ...ification Refocus and view the slide carefully Adjust the diaphragm again until the image has the best contrast Repeat with the 40x objective for 400x magnification Maintenance Adjusting the Stage Sto...

Page 5: ...eyepiece lenses are dirty Clean the lenses See Cleaning p 2 2 There is too much light Adjust the diaphragm Spots in field 1 The specimen slide objective or eyepiece lens is dirty Clean the slide or le...

Page 6: ...It was later known that the cells in cork are only empty because the living matter that once occupied them has died and left behind tiny pockets of air You can take a closer look at the cells also ca...

Page 7: ...Ltd 2006 Page 7 of 8 Visit us at ww homesciencetools com Date of slide Name of sample Collected from Stain Mount Lighting Observations Sketches 40x magnification 400x magnification Other ____________...

Page 8: ...perable due to excessive dirtiness as a result of misuse or lack of normal maintenance Any cameras and software supplied with this microscope are warranted for one year from the date of purchase You w...

Reviews: