background image

©

 Home Training Tools Ltd. 2012                                        

Page 7 of 8

                                         Visit www.homesciencetools.com

 

Pull bulb straight out

Tension collar

4.

 

Loosen the locking screw on the hinged 
door and swing the door open.  

5.

 

Gently grasp the halogen bulb. Pull it 
straight out of the lamp socket.  

6.

 

Use a tissue or cloth to grasp the new bulb. 
Insert it straight into the lamp socket, then 
close the door and tighten the locking screw.  

 
Adjusting Tension 

While the coarse focus 

tension is pre-adjusted by 
the manufacturer, if it falls 
out of adjustment, the stage 
will drift down under its own 
weight moving the image out 
of focus. To tighten tension, 
turn the collar clockwise.  

1.

 

Leave the tension as loose as possible. 
Tighten only enough to keep the stage from 
drifting downward.   

 
Replacing the Fuse 

If your microscope light flickers or fails to 

operate, the fuse may be burned out or not 
locked into position. To change the fuse, follow 
these steps:  

1.

 

Obtain the appropriate T1L 250 
volt replacement fuse.  

2.

 

Lay the microscope carefully on 
its side.  

3.

 

Unscrew the fuse holder on the bottom of 
the microscope.  

4.

 

Pull the fuse straight out from the socket and 
replace with a new one.  

5.

 

Screw the fuse holder back in.  

 

 

 
 
 
 

 
Troubleshooting 

If you experience difficulty with your microscope, try these troubleshooting techniques: 

 

 

Problem 

Possible Reason and Solution 

Light fails to 
operate 

1.

 

The AC power cord is not connected. 

Make sure the cord is fully connected to the 

socket on the microscope. Connect the cord to an outlet. 

2.

 

The bulb is burned out. 

Replace the bulb. (See ‘‘Changing the Bulb,’’ p. 7.) 

3.

 

The power source outlet is inoperative. 

Have a qualified electrician repair the outlet. 

4.

 

The incorrect bulb is installed. 

Replace with the correct bulb.

 

5.

 

Power switch is off or intensity control turned too low. 

Turn power on or rotate 

illumination intensity control.

Light flickers 

1.

 

The bulb is not properly inserted into the socket. 

Fully insert the bulb. 

2.

 

The bulb is about to burn out. 

Replace the bulb. 

3.

 

The connection at the AC outlet is loose. 

Have a qualified electrician repair the outlet.

4.

 

Fuse is not tightly screwed in.

 Screw fuse in all the way.

No image 

1.

 

The nosepiece is not indexed properly. 

Move revolving nosepiece until the objective

lens clicks into position. 

2.  The light is too bright. 

Adjust the diaphragm or illumination intensity control. 

Unable to 
focus slide 

1.

 

The slide coverslip is too thick. 

Use 0.17 mm thick (No. 1) coverslip. 

 

2.

 

The stage drops under its own weight. 

Adjust tension of coarse focus knob.

 

(See 

“Adjusting Tension,” p. 6.) 

 

3.

 

The slide is upside down. 

Place the slide on the stage with the coverslip facing up.  

 

Poor 
resolution,  
image not 
sharp 

1.

 

The condenser, objective, or eyepiece lenses are dirty. 

Clean the lenses. (See 

“Cleaning,” p. 2.) 

2.

 

There is too much light. 

Adjust the diaphragm or illumination intensity control.

 

Summary of Contents for MI-6000BIN

Page 1: ...RI Professional Laboratory Trinocular Microscope MI 6000DHD Professional Laboratory Dual Head Microscope MI 6000STD Professional Laboratory Microscope 1 800 860 6272 www homesciencetools com Copyright...

Page 2: ...al Camera 6 MI 6000STD Model 6 Maintenance 6 Changing the Bulb 6 Adjusting Tension 7 Replacing the Fuse 7 Troubleshooting 7 Specifications 8 Warranty 8 General Microscope Care Unpacking The Profession...

Page 3: ...al if you focus your specimen using one objective it will stay coarsely focused when you move to another objective you will still have to adjust the fine focus The 40x and 100x objectives are retracta...

Page 4: ...ocular head for enhanced image examination and extended viewing comfort Using the Binocular Head To best use this feature you must set the interpupillary distance to match the distance between your ey...

Page 5: ...using the stage adjustment controls Adjust the small fine focus knob until the specimen is clearly in focus 10 Adjust the iris diaphragm to reduce light by pulling the lever toward you when facing th...

Page 6: ...Model Trinocular Head The MI 6000TRI model comes with a trinocular teaching head With its comfort viewing 30 inclined binocular eyepieces and a vertical eyepiece it is ideal for simultaneous viewing...

Page 7: ...ason and Solution Light fails to operate 1 The AC power cord is not connected Make sure the cord is fully connected to the socket on the microscope Connect the cord to an outlet 2 The bulb is burned o...

Page 8: ...osition coarse and fine coaxial focusing controls with slip clutch tension adjustment and precise 0 002 mm fine focus division Focus Rack All metal rack and pinion focusing Stage Acid and chemical res...

Reviews: