Spots in field 

1.

 

The condenser, objective, or eyepiece lenses are dirty. 

Clean the lenses. (See 

“Cleaning,” p. 2.)

 

2.

 

The specimen slide is dirty. 

Clean the slide.

  

Uneven 
illumination of 
field 

1.

 

The nosepiece is not indexed properly. 

Move revolving nosepiece until the objective 

lens clicks into position.  

2.

 

The diaphragm is not properly indexed. 

Adjust the diaphragm to the proper level. 

Specifications 

Eyepiece 

Widefield 10x, large 20 mm diameter lens, with 20 mm exit pupil and fully coated optics.

Head 

Binocular head, 30° inclined Seidentopf with interpupillary adjustment and dual diopters
(MI-6000BIN).  
Trinocular, 30° inclined Seidentopf binocular eyepiece with interpupillary adjustment and 
dual diopters and vertical eyepiece with diopter (MI-6000TRI).  
Dual Head, 30° inclined eyepiece and vertical eyepiece with diopter (MI-6000DHD). 
Monocular, 30° inclined head rotates 360° (MI-6000STD). 

Nosepiece 

4-hole nosepiece is ball-bearing mounted with positive click stops. 

Objectives 

All plan objectives are DIN achromatic, parfocalled, parcentered, and fully coated.

 

4x, 0.10 N.A., red ring, 4.5 mm field of view, 40x magnification 

 

10x, 0.25 N.A., yellow ring, 1.8 mm field of view, 100x magnification 

 

40xR, 0.65 N.A., blue ring, 0.45 mm field of view, 400x magnification, retractable 

 

100xR, 1.25 N.A., white ring, 0.18 mm field of view, 1000x magnification, 
retractable, oil immersion  

Focusing 

Separate low position coarse and fine coaxial focusing controls with slip clutch, tension 
adjustment, and precise 0.002 mm fine focus division. 

Focus Rack 

All metal rack-and-pinion focusing.

Stage 

Acid and chemical resistant 132 x 140 mm metal stage with stage clips and integrated
mechanical stage. 

Mechanical 
Stage 

Low-position coaxial x-y controls provide precise 75 mm movement on the x-axis and 50
mm movement on the y-axis.  

Condenser 

Adjustable 1.25 N.A. Abbe condenser with rack-and-pinion focusing and centering 
controls. 

Diaphragm 

Iris diaphragm with 2 mm to 30 mm diameter opening and swing-out filter holder. 

Illuminator 

20-watt, 6-volt halogen illuminator with variable brightness. The illuminator works with 
100-240 volt AC power.  

  

Warranty 

Home Science Tools warrants this microscope to be free from defects in material and workmanship under 
normal use and service for the life of the instrument. Warranty does not cover light bulbs, batteries, or 
damage due to misuse, abuse, alterations, or accident. Warranty does not cover lenses that have 
become inoperable due to excessive dirtiness as a result of misuse or lack of normal maintenance.  

 

Any cameras and software supplied with this microscope are under warranty for one year from the 
purchase date.  

Return your microscope freight prepaid for warranty service to Home Science Tools, or the repair facility 
we designate. We will repair or replace your microscope at no charge and return it freight prepaid to you. 
Please call 1-800-860-6272 to arrange warranty service before returning this instrument. Please note that 
warranties apply only to the original purchaser and are not transferable.

 

Summary of Contents for MI-6000BIN

Page 1: ...RI Professional Laboratory Trinocular Microscope MI 6000DHD Professional Laboratory Dual Head Microscope MI 6000STD Professional Laboratory Microscope 1 800 860 6272 www homesciencetools com Copyright...

Page 2: ...al Camera 6 MI 6000STD Model 6 Maintenance 6 Changing the Bulb 6 Adjusting Tension 7 Replacing the Fuse 7 Troubleshooting 7 Specifications 8 Warranty 8 General Microscope Care Unpacking The Profession...

Page 3: ...al if you focus your specimen using one objective it will stay coarsely focused when you move to another objective you will still have to adjust the fine focus The 40x and 100x objectives are retracta...

Page 4: ...ocular head for enhanced image examination and extended viewing comfort Using the Binocular Head To best use this feature you must set the interpupillary distance to match the distance between your ey...

Page 5: ...using the stage adjustment controls Adjust the small fine focus knob until the specimen is clearly in focus 10 Adjust the iris diaphragm to reduce light by pulling the lever toward you when facing th...

Page 6: ...Model Trinocular Head The MI 6000TRI model comes with a trinocular teaching head With its comfort viewing 30 inclined binocular eyepieces and a vertical eyepiece it is ideal for simultaneous viewing...

Page 7: ...ason and Solution Light fails to operate 1 The AC power cord is not connected Make sure the cord is fully connected to the socket on the microscope Connect the cord to an outlet 2 The bulb is burned o...

Page 8: ...osition coarse and fine coaxial focusing controls with slip clutch tension adjustment and precise 0 002 mm fine focus division Focus Rack All metal rack and pinion focusing Stage Acid and chemical res...

Reviews: