AVERTISSEMENT: Comme condition de la certification, un connexion filaire ne peut pas être
faite via le port de communication lorsque le détecteur de gaz est dans la zone dangereuse à
moins que la zone soit connue pour ne pas être dangereuse.
WARNING: Li-Ion rechargeable battery: Only use approved battery pack: W01-3007-000 or W01-
3007-100.
AVERTISSEMENT: Batterie rechargeable Li-Ion: Utilisez uniquement des batteries approuvé:
W01-3007-000 ou W01-3007-100.
USE IN HAZARDOUS AREAS
Equipment which is intended for use in explosive atmospheres and which has been assessed
and certified according to international regulations may be used only under specified
conditions. The components may not be modified in any way.
i.
Connection must only be made to the Vdc input jack and to the USB communication port
when the instrument is in non-hazardous location or unless the area is known to be non-
hazardous.
ii.
The battery pack shall only be charged or replaced in a non-hazardous location.
iii.
The sensors shall only be replaced in a non-hazardous location.
iv.
The external connection to the PGM 65xxx shall be in accordance to control drawing W01
0901-CTL. The external alarms that are to be connected to PGM 65xxx must be suitable
for the hazardous location where they are mounted, and supplied from Class 2 power
supply or equivalent. The wiring method shall be to local electrical code and installation
are subject to acceptance by authority having jurisdiction.
The appropriate regulations for service and repair must be properly observed during such
activities.
PGM-65XXX contains a connector for mounting of a wind sensor – RAEMet. This connector has
a set of entity parameters that match the input entity parameters of RAEMet:
Uo: 5.2V; Po: 0.788W; Lo: 198
µ
H; Co: 1000
µ
F.
Keep all ports covered when they are not in use. This keeps moisture and debris out of the ports
and contributes to the instrument’s safety.
AreaRAE
11
User Manual
Summary of Contents for AreaRAE Plus
Page 1: ...AreaRAE WIRELESS MULTI GAS MULTI THREAT DETECTOR ...
Page 2: ......
Page 17: ...AreaRAE 16 User Manual ...
Page 22: ...Temperature Relative humidity AreaRAE 21 User Manual ...
Page 34: ...The instrument should wake up and begin monitoring immediately AreaRAE 33 User Manual ...
Page 145: ...Rev C September 2021 User Manual P N W01 4001 000 ...