If you have not made a change, it exits to the submenu’s next selection. If you have made a
change, you are prompted at the next screen to press [Y/+] to save the change or [N/-] to undo
the change and exit to the next submenu selection.
Zero at Start
If your AreaRAE Plus/Pro has been configured to perform a zero (fresh air) calibration upon
startup, called Zero At Start, then the startup routine is interrupted so that you can perform a
fresh air calibration for all sensors prior to using the instrument.
1.
Press [Y/+] to see on or off state, and which is selected.
2.
Press [N/-] to scroll to “On” or “Off.”
3.
Press [Y/+] to select.
4.
Press [Y/+] to “Save.” You can also press [N/-] to undo.
If you do not want to perform a zero calibration, press [MODE] to bypass it. If you start a zero
calibration and want to abort it, press [N/-], and the calibration stops and the main display is
shown.
Fast Startup
Fast Startup reduces the amount of time between when the instrument is turned on and is ready
for use. It skips showing you many settings and is best suited to environments where the
AreaRAE Plus/Pro is turned on and off very often during a given day. If Fast Startup is not
selected, then when the instrument starts, it shows you details of each sensor, including
calibration information, high and low alarm settings, etc.
1.
Press [Y/+] to see on or off state, and which is selected.
2.
Press [N/-] to scroll to “On” or “Off.”
3.
Press [Y/+] to select.
4.
Press [Y/+] to “Save.” You can also press [N/-] to undo.
Language
English is the default language, but other languages can also be selected for the instrument.
Note: The language can also be changed through Safety Suite Device Configurator (SSDC).
1.
Press [Y/+] to see the current, selected language.
2.
Press [N/-] to scroll to the language you want.
3.
Press [Y/+] to select.
4.
Press [Y/+] to “Save.” You can also press [N/-] to undo.
Site ID
Choose and enter an 8-digit Site ID to uniquely identify the particular site where the instrument
is to be used. The first four digits can be an alphabet letter or number, while the last four digits
can only be numbers. This Site ID is included in the datalog report.
AreaRAE
79
User Manual
Summary of Contents for AreaRAE Plus
Page 1: ...AreaRAE WIRELESS MULTI GAS MULTI THREAT DETECTOR ...
Page 2: ......
Page 17: ...AreaRAE 16 User Manual ...
Page 22: ...Temperature Relative humidity AreaRAE 21 User Manual ...
Page 34: ...The instrument should wake up and begin monitoring immediately AreaRAE 33 User Manual ...
Page 145: ...Rev C September 2021 User Manual P N W01 4001 000 ...