Enabling 3-Point Calibration for the VOC Sensor Via Safety Suite Device
Configurator (SSDC)
The AreaRAE Plus/Pro must be connected to a PC through the supplied USB cable and must be
in the PC Communications mode.
1.
Start up the Safety Suite Device Configurator (SSDC) software, enter a password, and
detect the instrument following the directions provided in the Safety Suite Device
Configurator (SSDC) User’s Manual.
2.
Click on the device's serial number to view the AreaRAE Plus/Pro’s current configuration
information.
3.
Click on the Sensors tab.
4.
Scroll down until VOC sensor is visible.
5.
Click on the edit icon
.
6.
Click on the checkbox to activate 3-Point Calibration
7.
Click the Save button
8.
When you are done, quit Safety Suite Device Configurator (SSDC) and then press [Y/+] on
the AreaRAE Plus/Pro to exit the PC communications mode. The instrument returns to
operating in Normal mode.