BW Clip Real Time

2 OR 3 YEAR H

2

S, CO, O

2

, SO

2

 SINGLE-GAS DETECTOR

Activate a New Detector

1.  Move to a normal atmosphere (20.9% v/v O

2

) that is free of hazardous gas.

2.  Press and hold the button until a 5 second countdown is displayed, and then 

continue to hold until the countdown is complete.

3.  When the countdown is complete, the LCD and LEDs turn on and then turn off. The 

detector performs an internal diagnostic test.

4.  When the diagnostic test is successful, the alarm setpoints are displayed and the 

LEDs flash.

5.  The sensor stabilization countdown is displayed. During the countdown, the detector 

vibrates continuously for 20 seconds. The time required to stabilize varies depending 
on sensor type. When the countdown reaches 0, the LEDs flash and the audible alarm 
beeps.

6.  The detector is in normal operating mode when the gas type and concentration are 

displayed.

Normal Operating Mode

When the detector is in normal operating mode, the type of gas 
detected is permanently displayed. The detected concentration of the 
gas is displayed until it is disrupted by a button action, gas alarm, or 
error event.

Operating Life

When the detector is activated for the first time, the operating life 
countdown is displayed. Press the pushbutton to display remaining 
operating life.

90 Days or Less Remaining

When the operating life countdown reaches 90 days or less before 
expiry, the countdown display changes to days. The countdown 
displays the remaining operating life in days until the countdown 
reaches 24 hours or less before expiry.

24 Hours or Less Remaining

When the operating life countdown reaches 24 hours or less before 
expiry, the countdown display changes to hours, and the 

EXPIRY!

 

warning and non-compliance symbol are displaye

d.

End of Operating Life

When the operating life countdown ends, the detector deactivates 
and detector safety functions are disabled. It is possible to retrieve 
event logs for a limited time after expiry. The 

EXPIRY!

 warning and 

non-compliance symbol are displayed for up to 30 days after expiry.

Internal Diagnostic Tests

Activated detectors automatically perform one internal diagnostic test every 24 hours. If 
the internal diagnostic test fails, the diagnostic fail-safe begins.

Diagnostic Fail-Safe

1.  The 

EXPIRY!

 warning and non-compliance symbol are displayed.  

2.  The detector beeps and vibrates.
3.  The LEDs flash 2 times per second for 15 seconds. 
4. 

OFF

 or an error code is displayed. The detector is deactivated. 

It is possible to retrieve the event logs for a limited time after expiry. If you are unable to 
retrieve the event logs, contact Honeywell.

Alarms and Warnings

An alarm is initiated when the sensor is exposed to a gas concentration that exceeds 
alarm setpoints. The alarm persists until the alarm gas concentration returns to an 
acceptable range. Battery life decreases rapidly when the detector is in alarm condition. 

Low Alarm

Audible

: 1 beep per second

Visual:

 1 flash per second

Vibration:

 1 vibration per second

High Alarm

Audible:

 2 beeps per second

Visual:

 2 flashes per second

Vibration:

 2 vibrations per second

Over-Limit and Under-Limit Alarms

If the gas concentration is beyond the sensor range, 

OL

 (over-limit) 

or 

-OL

 (under-limit) is displayed.

Audible:

 2 beeps per second

Visual:

 2 flashes per second

Vibration:

 2 vibrations per second

Automatic Zero Reminder for O

2

 Models

For O

2

 models only, the 

Automatic Zero Reminder

 is displayed when 

more than 24 hours have elapsed since the last successful zero 
procedure was performed.

Peak Readings

The peak readings symbol is displayed when the sensor is exposed to 
a gas concentration that exceeds the alarm setpoints. It is no longer 
displayed when more than 24 hours have passed since the last alarm, 
or when a successful bump test is performed via an IntelliDoX station.

Non-compliance Symbol and LED Indicators

The non-compliance warning symbol is displayed when:

•  an internal diagnostic test fails;
•  a bump test, calibration or zero procedure fails;
•  a bump test or calibration is due, if enabled;
•  the sensor is exposed to a gas concentration that exceeds alarm setpoints, if 

enabled.

The non-compliance warning symbol is no longer displayed when more than 24 hours 
have passed since the last alarm, or when a successful bump test or calibration is 
performed (manual calibration or via an IntelliDoX station). The non-compliance warning 
symbol is permanently displayed when the service-life countdown reaches 24 hours or 
less before expiry, and for up to 30 days after expiry.

When 

Non-compliance after gas exposure

 is enabled through Safety Suite Device 

Configurator via an IntelliDoX station, the LEDs flash in an alternating pattern when the  
non-compliance warning symbol is displayed. When the symbol is first displayed, the top 
LEDs flash. After 30 seconds, the side LEDs flash. After 30 seconds, the pattern repeats 
until the non-compliance warning symbol is no longer displayed.

NOTE: 

When the detector is bump tested via an IntelliDoX station, an abnormal test 

result may occur if non-compliance LEDs are flashing. To suspend the flashing for up to 
10 minutes, press and release the button. When the flashing stops, insert the detector 
and perform the test. If the bump test fails, the LEDs resume flashing.

Recall Events and Settings

Note: 

Alarm setpoints can be manually setup via an IntelliDoX device.

When the detector is in normal operating mode, press the button at any time to scroll 
through events and settings that are not empty or disabled through Safety Suite Device 
Configurator via an IntelliDoX station. BW Clip Real Time flashes, vibrates and beeps.
While an event or setting is displayed, press the button to scroll to the next available 
event or setting. If you do not press the button, or if you have reached the last available 
event or setting, the detector returns to normal operation. Events and settings are 
displayed in this order:
1. 

Peak Reading

, if available, and when a maximum gas exposure event has occurred 

within the past 24 hours.

2. 

Time of Peak Reading

, if Real-time Clock display is enabled.

3. 

Operating Life

, the remaining operating life of the detector in months, days or hours.

4. 

Low Alarm Setpoint

5. 

High Alarm Setpoint

6. 

Real-time Clock, 

if enabled.

7. 

Firmware Version

8. 

Next Bump Due

, if enabled.

9. 

Next Calibration Due

, if enabled.

10. 

Hibernation

, for 2-year H

2

S and CO models only.

Summary of Contents for BW Clip Real Time

Page 1: ... BW Clip Factory Calibration Certificate This instrument was factory inspected tested and calibrated in accordance with the conditions and requirements of our registered Quality System Operating Standards and Sales Agreements MISE EN GARDE 1 Activez le détecteur avant la date d activation sur l emballage 2 Ce produit est un détecteur de gaz pas un dispositif de mesure 3 Veillez à assurer que la gr...

Page 2: ...s per second Over Limit and Under Limit Alarms If the gas concentration is beyond the sensor range OL over limit or OL under limit is displayed Audible 2 beeps per second Visual 2 flashes per second Vibration 2 vibrations per second Automatic Zero Reminder for O2 Models For O2 models only the Automatic Zero Reminder is displayed when more than 24 hours have elapsed since the last successful zero p...

Page 3: ... IntelliDoX station to enable the following optional settings for BW Clip Real Time detectors Non compliance After Gas Exposure When Non compliance after gas exposure is enabled the LEDs flash in an alternating pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints When the gas exposure is detected the top LEDs flash After 30 seconds the side LEDs flash After 30 sec...

Page 4: ...afe area that is free of hazardous gas Recommended Calibration Gas Mixtures H2 S 5 to 50 ppm balance N2 CO 25 to 250 ppm balance N2 SO2 5 to 70 ppm balance N2 O2 5 0 to 19 0 v v O2 Perform the Zero procedure once every 24 hours or when the Automatic Zero Reminder is displayed Intended Use This product is classified for use in hazardous atmospheres that are not more than 21 v v O2 Recycling This in...

Reviews: