Error Codes
Type
Code
Display message
Key
Action
Note
Forbidden
Error
5001 Battery Critical Low
Powering
off…
Power off in 3 seconds after this
message
5002 Battery Too Hot
Powering
off…
Power off in 3 seconds after this
message
5003
Battery Module No
Response
Powering
off…
Power off in 3 seconds after this
message
Fatal Error 4002
Contact your
distributor
Hold to
power off
Power off
in 300s
Unknown product model
4003
Contact your
distributor
Hold to
power off
Power off
in 300s
Application instrument ID set for
service mode
4004
Contact your
distributor
Hold to
power off
Power off
in 300s
Sensor instrument ID does match
application module
4005
Contact your
distributor
Hold to
power off
Power off
in 300s
Sensor instrument ID set for service
mode
4006
Contact your
distributor
Hold to
power off
Power off
in 300s
No response from sensor module
4008
Contact your
distributor
Hold to
power off
Power off
in 300s
Incompatible sensor and
application firmware
4011
Contact your
distributor
Hold to
power off
Power off
in 300s
Bad battery (Battery voltage too
high or too low during charging)
4013
Contact your
distributor
Hold to
power off
Power off
in 300s
No battery (Power supplied from AC
adapter)
4015
Contact your
distributor
Hold to
power off
Power off
in 300s
Lost communication
4017
Contact your
distributor
Hold to
power off
Power off
in 300s
Input wrong password 5 times
Critical
Error
3001 RTC Error
Click to
acknowledge
Power off
in 300s
3003 Pump not found
Click to
acknowledge
Power off
in 300s
Reinstall Pump
3005
Gas plate not
installed
Click to
acknowledge
Power off
in 300s
Reinstall Gas Plate
Normal
Error
2001
Failed to Turn on
Lamp
Click to
acknowledge
Power off
in 300s
Cannot turn on PID lamp
Message
1001 Datalog Link Broken
Click to
acknowledge
Power off
in 300s
Datalog is broken
1005
Unsupported Sensor
Found and
Deactivated
Click to
acknowledge
Power off
in 300s
Check sensor installed. See
"Replacing Sensors" on page 83 for
more information.
BW RigRat
97
User Manual
Summary of Contents for BW RigRat
Page 1: ...Honeywell BW RigRat LOCAL AREA GAS MONITOR ...
Page 2: ......
Page 6: ......
Page 107: ...User Manual P N W03 4001 000 BW RigRat Revision E Honeywell September 2022 ...