Metering Lines (4)
6 - 6
Model 1010 Application Pack CB
Part No.: 4418024 - Rev. 1
Programming Manual
4 mA s 212 °F (for US units)
Or
4 mA setting to Max range in Table 6-1
Meter n
Min Error Limit
The minimum allowed temperature range. A temperature below this
value flags an error when loading.
Constraints:
Only available when you use the Weights and Measures switch on
the instrument to enter programming mode.
Only if Temp. Sensor = RTD
Enter a value in the following range:
Refer to Table 6-1 on page 6-5.
NOTE:
Use the Arm key to toggle between negative and positive values.
Meter n
Max Error Limit
The maximum allowed temperature range. A temperature above this
value flags an error when loading.
Constraints:
Only available when you use the Weights and Measures switch on
the instrument to enter programming mode.
Only if Temp. Sensor = RTD
Select one of the following options:
Min Error Limit to Max range in Table 6-1 on page 6-5.
Breakdown
Temperature
Allows enabling or disabling the breakdown temperature.
Constraints:
Only available when you use the Weights and Measures switch on
the instrument to enter programming mode.
Only if temperature correction is enabled.
Select one of the following options:
Enable
Disable*
NOTE:
When the break down value is enabled the unit no longer meets the MID
standard.
Meter n
Breakdown
Temperature
Allows entering the breakdown temperature. This value is used for
correction calculations if the temperature probe fails.
Constraints:
Only available when you use the Weights and Measures switch on
the instrument to enter programming mode.
Only if temperature correction and breakdown temperature are
enabled.