Calibration
Part No.: 44200004 - Revision 2
SVP Controller Operation Manual
4-9
21. Calculate prover volume per Section 4.5.
22. Enter the average calibration volume calculated in the previous step
by selecting the “Enter Average Volume” menu. This screen can
be accessed while still in the Calibration mode or at a later time by
entering the Calibration menu and then accessing this screen directly.
After selecting Enter Average Volume following screen appears
Enter Average Volume
5.37000
7
8
9
Backspace
4
5
6
Confirm
1
2
3
0
-
.
Enter value
Stop
23. After the Average Volume has been entered and confirmed the
following screen is displayed to enter the Repeatability in percent (%).
The entered average volume and the repeatability values will be stored
in the log with a timestamp of the entry.
After selecting Enter Average Volume following screen appears
Enter Repeatability
0.00200 %
7
8
9
Backspace
4
5
6
Confirm
1
2
3
0
-
.
Enter value
Stop
Selecting “Abort Calibration” will stop the calibration process. This action
will stop the motor if in running state, bypass value (SV2) will be turned ON
and draw valve (SV1) will be turned OFF. If the Prover is in sweep mode
with the piston moving downstream no action will be taken for upstream
and downstream sensor inputs. Calibration log entry will not be made.