43
GasAlertMicro 5/PID/IR
Alarms
To change the factory-set alarm setpoints, refer to
Calibration and Set-
ting Alarm Setpoints
.
Note
To disable an alarm, set the alarm setpoint to
0
(zero).
Stopping a Gas Alarm
The low and high alarms stop when the ambient gas level returns to
below the low alarm setpoint.
Note
If alarms are set to latch, press
C
to reset the alarms.
a
Caution
Follow all safety procedures as defined by your
employer. Confirm with your supervisor before clearing
TWA and STEL alarms.
The TWA and STEL alarms can be stopped either by
• clearing the MAX, TWA, and STEL exposures (refer to
Clearing
Gas Exposures
), or
• deactivating the detector for 5 minutes (minimum) and then
reactivating it again.
Sensor Alarm
The detector tests for missing or defective sensors during the activation
self-test. If a sensor fails the self-test,
FAIL
flashes on the LCD above
the failed sensor. Refer to
Troubleshooting
.
Pump Alarm
The external pump draws air over the sensors continually. If the pump
stops operating or becomes blocked, the detector activates the pump
alarm. The following screens display.
The pump alarm continues until the blockage is cleared or it is
acknowledged by pressing
C
. If
C
is pressed, the detector automati-
cally launches a pump test to reset the pump module.
Summary of Contents for GasAlertMicro 5IR
Page 1: ......
Page 8: ...GasAlertMicro 5 PID IR User Manual Title Page ...
Page 10: ...GasAlertMicro 5 PID IR User Manual ...
Page 12: ...GasAlertMicro 5 PID IR User Manual ...
Page 104: ...GasAlertMicro 5 PID IR User Manual ...
Page 105: ...iERP 128900 D5615 5 English BW Technologies 2009 All rights reserved ...