GasAlertMicro 5/PID/IR
User Manual
Table 21. Troubleshooting
Problem
Possible Cause
Solution
Detector Operation Troubleshooting
Detector does not display
normal ambient gas
readings after startup
self-test.
Sensor not stabilized
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Detector requires calibration
Calibrate the sensors. Refer to
Calibration and Setting Alarm Setpoints
.
Target gas is present
Detector is operating properly. Use caution in suspect areas.
Detector does not respond
to pushbuttons.
Batteries are depleted
Refer to
Replacing the Alkaline Batteries
or
Replacing the Lithium Battery
Pack
.
Detector is performing
operations that do not require
user input
Pushbutton function restores automatically when the operation ends.
Detector does not
accurately measure gas.
Detector requires calibration
Calibrate the sensors. Refer to
Calibration and Setting Alarm Setpoints
.
Detector is colder/hotter than
ambient gas
Allow the detector to adjust to ambient temperature before using.
Sensor filter is blocked
Clean the sensor filter. Refer to
Replacing a Sensor or Sensor Filter
.