GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
P.4510EN.01
12/5/2007
26
Access Level: General
You can use this menu to set the
initial
NIBP cuff inflation setting on a single-user monitor. (Otherwise,
set the initial NIBP on the patient ID card.)
Genesis DM uses this initial setting only for the
first
blood pressure collection because the monitor
stores the latest BP reading in memory to use as a setting for the next blood pressure. The monitor
inflates the cuff to 35 mmHg higher than the last systolic reading.
Once the monitor is powered down the default setting of 175 mmHg is re-established.
In the
System Settings
menu, scroll to
Inflation Settings
, and press
RETEST
to enter. The current
setting blinks.
Press either
YES
or
NO
to scroll to the desired inflation setting, and press
RETEST
to accept the
new setting.
You can scroll to the next
System Settings
option -or- press
MANUAL
to return to the
Main
menu, or press
START + YES
to exit Setup Mode.
Display Brightness
Options:
Low, Med, High
Default:
High
Access Level: General
This menu allows you to adjust the brightness level of the VFD display. To adjust the brightness level,
follow the steps below.
1.
In
System Settings
, scroll to
Display Brightness
, and press
RETEST
to enter. The current
selection blinks.
2.
Press either
YES
or
NO
to scroll to the desired brightness, and press
RETEST
to accept the new
setting.
You can scroll to the next
System Settings
option -or- press
MANUAL
to return to the
Main
menu, or press
START + YES
to exit Setup Mode.
Questions Menu
The
Questions Menu
allows you to program the monitor to ask subjective disease-state questions
similar to the questions you might ask during a patient visit. There is a block of fifty-one standard