GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
P.4510EN.01
12/5/2007
30
Transmit Menu
In Setup Mode, use
YES
or
NO
to scroll to the
Transmit
Menu
, and press
RETEST
to enter. The
first option, PIN, displays.
PIN Number
Options: 7-digit
number
Default:
No default. Each unit has a unique PIN
Access Level: Password protected
The PIN number identifies your organization’s mailbox where monitor vitals and configurations are
stored until retrieved by Honeywell HomMed. This PIN is provided by Honeywell HomMed. The
Genesis DM monitor sends data to this mailbox address and the LifeStream™ checks this same address
to retrieve the data.
You must
make sure that this number is correct
. Check the number against the one you received
during your inservice.
1.
In the
Transmit
Menu
, scroll to the
PIN Number
option. If the number is correct, you can move to
your next task. However, if it is incorrect, press
RETEST
to enter the option. The first digit blinks.
2.
Use
YES or NO
to scroll to the correct number, and press
RETEST
to accept the value and
move to the next digit.
3.
Repeat steps 2 and 3 until you enter or correct the entire PIN number.
You can scroll to the next Transmit Menu option, press
MANUAL
to return to the Main menu, or
press
START + YES
to exit Setup Mode.
If you just need to correct one digit, use the
RETEST
key instead of the
RETEST
key. The
MANUAL
key will
enter the new digit and also back you out of the menu.
Dial Type
Options:
Tone or pulse (rotary)
Default:
Tone
Access Level: General
If tone dialing is unavailable and there is a rotary phone, change the dial type to pulse.