GENESIS DM CLINICAL GUIDE
Patient Training
P.4510EN.01
12/5/2007
55
1.
Put 2-3 drops K2 solution on the external electrode and on the right thumb electrode.
2.
Hold the ecg@home device in the right hand with the thumb gently pressing the electrode.
3.
Place the external electrode on the upper left leg.
4.
With the left hand, press START to turn the unit on and then gently hold the external electrode in
place while the device takes a reading.
5.
When the upload is complete, the monitor displays the text, “Data Upload finished.”
Additional notes
•
The patient should attempt at least two times to obtain a reading without noise. Even if the second
reading has noise, the patient should still transmit the reading.
•
The ecg@home does not have an OFF button. The device automatically turns off after five minutes.
When connected to the Genesis DM monitor, the monitor turns the device off after a successful
transmission.
•
The monitor erases the reading prior to turning the device off.
•
If the patient uses the device as a stand-alone (not connecting it to the monitor), the device erases
the old reading when the patient turns it on.
Cleaning the ecg@home
Always disconnect the device from the Genesis DM monitor prior to cleaning. Clean the device with a
soft cloth moistened with the K2 solution. Do not use water, detergent or any other cleaning product.
To clean the electrodes, moisten a cotton swab with the K2 solution and gently swab clean. To
disinfect
the unit, use a solution of isopropyl alcohol (60-70%) and water (30-40%).