GENESIS DM CLINICIAN GUIDE
Getting Started
P.4510EN.01
12/5/2007
5
WARNING
EXPLOSION HAZARD
This unit is powered with electricity and could ignite highly flammable gases.
DO NOT use this device in the presence of explosive or flammable
anesthetic agents.
WARNING
NOT A NEONATAL UNIT
Using the Genesis DM monitor on infants or small children could cause serious injury.
DO NOT use the monitor, temperature probe, blood pressure cuff, or any
monitor accessory on infants or small children.
WARNING
NOT DEFIBRILLATOR PROOF
Using the Genesis DM monitor on patients undergoing defibrillation may damage the
monitor or peripheral equipment.
DO NOT use the monitor on patients undergoing defibrillation.
WARNING
PHYSICAL HAZARD
Please exercise care in the routing of external cables, wires and hoses to avoid creating
a tripping hazard.
DO NOT route cables across a walkway
CAUTION
SENSITIVE EQUIPMENT
The monitor contains sensitive components. Liquids or chemicals will seriously damage
the unit.
DO NOT autoclave or ethylene oxide sterilize.
DO NOT immerse any equipment in liquid.
If liquid is spilled on the monitor, or if the unit has been dropped or damaged, contact
Honeywell HomMed Customer Service.