14
recognized. At the top of the right side of
the screen there is a tab for each dock
that is currently installed and recognized.
Within each dock tab are two sub-tabs:
one for dock controls and one that shows
the calibration status and details for the
instrument that is currently in the dock.
3.1 Dock Status (USB Connection)
Note: If the dock is connected to the
PC via Ethernet, see the Status
Indicators section of the IQ Ethernet
Instructions Booklet.
The current status of each dock is shown
in the left column. When there are no
detectors in the dock, each of the status
indicators will be shown in grey.
When a detector is placed in the dock, the
indicator will change to reflect the new
status. Status changes are also be shown
on the dock’s LCD.
During processing, the Processing
Indicator is shown in yellow.
The operations being performed on the
instrument are listed as they occur.
If the dock is able to successfully
complete the processing dictated by the
instrument’s template in the IQ Database
Manager Pro Program, OK will be shown
in green.
For more information regarding
template settings, see the IQ Database
Manager Pro Reference Manual.
If a fault is detected and the dock is
unable to complete the tests, the red fault
light will be shown on the screen. Details
of the fault will be listed in the input box on
the right side of the screen under the
instrument tab. See section 3.2.2 for
details on faults.
3.2 Dock Controls and Instrument
Status
The right side of the screen contains
individual controls for each MultiPro IQ
Express Dock. Each dock has its own
page, which can be accessed by clicking
on the appropriate tab at the top of the
page. Within each dock’s page are two
sub-tabs. One contains specific
information on the dock itself. The other
displays information on the instrument that
is currently recognized in the dock.
Note: If the “Change Config” button is
enabled, press it before making any
changes to the settings.
3.2.1 Dock Configuration
Click on the Dock tab to access the dock
settings.