Messages and Diagnostic Codes
Release 3
34-ST-25-20 MC Toolkit User Manual
65
9/06
Messages and Diagnostic Codes
Messages and Diagnostic Codes
Table 18 MC Toolkit/SDC 625 Error Messages
MC TOOLKIT/SDC 625 ERROR MESSAGES
MESSAGE CORRECTIVE
ACTION
Another Secondary master has been
detected! HART Protocol does not
support more than one secondary
master on the HART Network.
Remove one of the secondary masters from the network.
Checksum error on Modem Response!
A noisy environment can cause this error. Repeat the command again.
Com Port Read Timeout!
Ensure that the Honeywell supplied modem cable is used and that the
connections are secure.
Make sure the modem has a good battery.
Com read Error!
Error writing to Com Port!
Error writing to Com Port!
EscapeCom function Error!
GetCommState Error!
Make sure ActiveSync is not running.
If several programs are active, try closing one or more open programs.
Stop the MC Toolkit application by doing File | Exit and restart the
program.
Use the Pocket PC hardware reset. See your Pocket PC documentation
reset the unit.
Caution for MCT101 Users:
If the Symbol PPT 2800
computer is reset, it will lose all its installed programs and data. The
program will have to be installed again.
MCT101 Users, make sure you are using a Honeywell approved Pocket
PC.
Host Failed to get good response from
the Modem!
MCT101 Users: ensure that the Honeywell supplied modem cable is
used and that the connections are secure.
Make sure the modem has a good battery.
Invalid Command number from the
Modem!
Communications problem between the Pocket PC and the modem.
Repeat the command again.
Invalid Handle Value!
Make sure ActiveSync is not running.
If Several programs are active, try closing one or more open programs.
Stop the MC Toolkit application by doing File | Exit and restart the
program.
Use the Pocket PC hardware reset. See your Pocket PC documentation
reset the unit.
Caution for MCT101 Users:
If the Symbol PPT 2800
computer is reset, it will lose all its installed programs and data. The
program will have to be installed again.
Make sure you are using a Honeywell approved Pocket PC.