Messages and Diagnostic Codes
Release 3
34-ST-25-20 MC Toolkit User Manual
69
9/06
DE MESSAGES
Are you sure you wand to Reset All
Corrects to factory defaults?
Executing Reset All Corrects will overwrite all user input calibration
values (Zero, LRV, URV) with factory-default values.
Are you sure you want to place the
transmitter in output mode (?)
In Output Mode, output current will be set to the selected constant
value, rather than to calculated
Square Root is not allowed for GP and
AP Sensor Types.
Square Root conformity is available only for FLOW (DP) input
applications.
Table 20 HART Messages
HART MESSAGES
Bad Manufacturer Code
MC Toolkit does not recognize the manufacturer code from the
transmitter.
Bad Status code from the transmitter
MC Toolkit does not recognize the status code from the
transmitter.
Bad Start character on HART Message!
First byte in HART response message is not a valid byte(valid
bytes:0x02,0x82,0x06,0x86,0x01,0x81)
Device specific command error
Possible error codes 6, 8, 9, 10, 11, 12, 13, 14 and 15.
Subset of these is defined for commands. If any value other than
these subset values, this message is displayed.
Error code 02: Invalid Selection
The selection is not valid.
Error code 03: Passed parameter too large
Entered value is too large.
Error code 04: Passed parameter too small
Entered value is too small.
Error code 05: Too few data bytes received
Internal error. Repeat command.
Error code 06: Wrong Password!
The password entered is incorrect.
Error code 07: In Write Protect Mode
The transmitter is in Write Protect Mode.
Error Code 08: Set to Nearest Possible Value
Set to Nearest Possible Value
Error Code 09: Applied process too high
Applied process too high or out of range.
Error code 09: LRV too high
LRV is too high or out of range.
Error Code 09: Not in Proper Current Mode
Not in Proper Current Mode
Error Code 10: Applied process too low
Applied process too low or out of range.
Error code 10: LRV too low
LRV too low or out of range.
Error Code 11: Excess correction attempted
Attempted correction value is out of range.
Error Code 11: Invalid Transmitter variable
Selected variable is invalid for this transmitter.
Error Code 11: Transmitter in Multidrop Mode!
Action could not be completed because the transmitter is in
multidrop mode.
Error code 11: URV too high
URV too high or out of range.
Error Code 12: Invalid Units
Invalid Units