4. On the Result window the status of zero calibration is displayed. Click Finish.
5. In the Inhibit window you have the option to check Inhibit, and then click Accept.
Span Calibration
Perform an span calibration for each channel of the device.
1. On the home page, select Calibration > Span Calibration.
2. On the Span calibration page, you can select the gas, the gas type, and the target
concentration.
3. Click Start and then follow screen instructions.