Number
Date
Data Field
Condition
Recovery
F10
Baseline
Fault
Concentration
without
Deadband
Cell baseline has drifted.
Check for background
gas concentration,
temperature or
humidity fluctuations.
Perform zero
calibration. Replace
cell.
F20
Calibration
Expired too
long.
N/A
Too long since last
calibration.
Replace or calibrate
the Cell.
F21
Cell expired.
N/A
Cell is too old.
Replace the cartridge
or cell.
F40
Temperature
Limits
Exceeded
Temperature
(C)
Temperature is out of
limit.
Check installation
environment.
F41
Flow Fail
Flow (cc/min)
Flow < 70% of nominal
for 24 seconds
Check filters and
pump.
F42
Pump
Memory
error
N/A
Pump memory has no
response or corrupted
data
Reboot
F43
Pump
absent
N/A
There is no pump
Device Power off and
then insert the pump
F44
LIT test fail
N/A
LIT test failed
Check a flow path
F50
Cell Failure
N/A
Reflex test fail of toxic
cell
Replace cell
F51
Cell comm.
Failure
N/A
It has failed to
communicate with IR
cell
Replace IR cell
F52
Cell memory
error
N/A
Cell memory has no
response or corrupted
data
Replace cell
F53
Cell Memory
Invalid
N/A
Checksum error.
Replace cell
F54
Cell IR
N/A
Cell IR voltage is out of
Replace cell
Midas-M
71
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...