Number
Date
Data Field
Condition
Recovery
voltage error
range
F60
The absent
cartridge
N/A
No communications.
Reset cartridge.
Replace cartridge.
F61
Cartridge
wrong type
Cell ID
Cartridge has one or
more empty channel
Replace cartridge.
F62
No valid cell
N/A
Cartridge has no valid
cell
Replace cartridge.
F63
Cartridge
Memory
Invalid
N/A
Checksum error.
Reboot
F64
Cartridge
app Memory
Invalid
N/A
Checksum error.
Replace cartridge.
F65
Cartridge
3.3V error
Voltage
Cartridge 3.3V is out of
range
Replace cartridge.
F66
Cartridge 5V
error
Voltage
Cartridge 5V is out of
range
Replace cartridge.
F67
Cartridge
app failed
N/A
Cannot start cartridge
app
Update the latest
cartridge app
F80
Transmitter
NOR flash
error
N/A
No communication
Replace the
transmitter
F81
Transmitter
memory
invalid
N/A
Checksum error.
Reboot
F82
Transmitter
memory
comm. Error
N/A
Communication failed
Reboot
F83
Transmitter
app memory
invalid
N/A
Checksum error.
Replace transmitter or
update the latest
version
F84
Transmitter
3.3V error
Voltage
Transmitter 3.3V is out
of range
Replace Transmitter
Midas-M
72
User Manual
Summary of Contents for Midas-M
Page 2: ......
Page 6: ......
Page 9: ...User Interface Detector s User Interface overview Midas M 9 User Manual ...
Page 16: ...Midas M 16 User Manual ...
Page 25: ...Generic example Midas M Isolated 4 20mA output w PoE power Midas M 25 User Manual ...
Page 37: ...Option Operations you can Test from the selected option l Inhibit time Midas M 37 User Manual ...
Page 55: ...3 Wait until the gas reading is stable and then click Accept Midas M 55 User Manual ...
Page 63: ...Midas M 63 User Manual ...
Page 64: ......
Page 68: ......
Page 81: ......
Page 100: ...Midas M 100 User Manual ...
Page 104: ...The certificate should be reinstalled when the IP address is changed Midas M 104 User Manual ...
Page 109: ... 2019 ...