background image

Inserting Sensors

 

Important!

 Always make sure your hands are clean and dry before handling sensors. 

Place a sensor into a socket, aligning it with the matching keyways and the multi-pin connector at the bottom. Once a 
sensor is plugged in, its corresponding green LED flashes once and if it requires warm-up, then warm-up begins. Below 
are approximate times for sensor warm-up. 

Non-EC sensor: 

0 hours (no warm-up required)

 

Non-biased EC sensor:

 6 hours 

Biased EC sensor:

 24 hours 

If the green LED does not blink, the sensor is not inserted properly. 

Note:

 If the sensor is properly seated and you remove it at any time during its warm-up, it must be returned to its socket 

within 20 seconds, or the SensorRAE 4R+ will restart the warm-up cycle from the beginning. 

 

Removing Sensors 

Always lift sensors out with your fingers, gripping around the sides. 

Checking Sensor Status 

You can check the status of a sensor being warmed up by pressing the power button. If the green LED next to the 
sensor glows, it is fully conditioned and ready for use. 