12.4 Calibration Procedure
If you are not familiar with operating the instrument in the SETUP menu mode,
see “SETUP Menu mode” on page 23. A listing of all menu items is given in
“Setup Menu Reference” on page 133.
Step 1: Wire the transducer to the channel’s connector.
See “Wiring” on page 92 for details.
Step 2: Set the Excitation and Signal jumpers appropriate for the transducer,
amplifier or DC-DC LVDT.
See “Wiring” on page 92 and “Excitation and Signal Jumpers” on page 100.
Step 3: Enter the
CALIBRATION TYPE
.
There are two methods that can be used to calibrate the transducer to the
Input Channel. Each has advantages and disadvantages as described in
“
Menu Item” on page 106. It is important to know your
application in order to select the appropriate calibration type.
Step 4: Enter the
CALIBRATION DATA.
Otherwise, consult the Certificate of Calibration for the transducer when
entering information in the
CALIBRATION DATA
sub-menu.
Step 5: Perform the calibration.
Using the
CALIBRATE
menu item starts the calibration process. You will be
prompted to apply loads to the transducer as required.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...