compensate for the non-linearity in the transducer. The usual loads used are
zero scale, half scale and full scale, but you are not restricted to these loads.
•
“
TYPE= 5 POINT CAL
” means 5-Point Known Load Calibration. You are prompted
to apply the loads to the transducer that were entered in the “
KNOWN POINT 1/5
”,
“
KNOWN POINT 2/5
”, “
KNOWN POINT 3/5
”, “
KNOWN POINT 4/5
” and “
KNOWN POINT 5/5
” regis-
ters. This technique can be used to compensate for the non-linearity in the
transducer. The usual loads used are zero scale, one-quarter scale, half
scale, three-quarters scale and full scale, but you are not restricted to these
loads.
The selection of the calibration type affects what menu items appear in the
CALI-
BRATION DATA
sub-menu.