page 110 008-0671-00
12.6.6
CALIBRATE
Menu
Item
This menu item performs a calibration according to what was entered in the
CALI-
BRATION TYPE
and
CALIBRATION DATA
menu items.
If the
CALIBRATION TYPE
is
Shunt Calibration
...
•
The display will read
DOING SHUNT CAL
, and prompt you to
APPLY 00000 UNIT
(where
00000 and UNIT are previously entered as
ZERO SCALE VALUE
and
DISPLAY. UNITS
.
When you have applied this load to the transducer, press [ENTER].
•
The display will then read
WORKING
and calibration will take place. Upon com-
pletion, the display will indicate
DONE
and the instrument will return to the RUN
mode.
If the
CALIBRATION TYPE
is
2-Point Known Load Calibration
...
•
The display will read
DOING 2POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/2
and
DISPLAY.
UNITS
). When you have applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are
previously entered as
KNOWN POINT 2/2
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
Upon completion, the display will indicate
DONE
and the instrument will return to
the RUN mode.
If the
CALIBRATION TYPE
is
3-Point Known Load Calibration
...
•
The display will read
DOING 3POINT CAL
, and prompt you to
APPLY 00000 UNIT
(where 00000 and UNIT are previously entered as
KNOWN POINT 1/3
and
DISPLAY.
UNITS
). When you have applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 00500 UNIT
(where 00500 and UNIT are
previously entered as
KNOWN POINT 2/3
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
The display will read
WORKING
, then
APPLY 01000 UNIT
(where 01000 and UNIT are
previously entered as
KNOWN POINT 3/3
and
DISPLAY. UNITS
). When you have
applied this load to the transducer, press [ENTER].
•
Upon completion, the display will indicate
DONE
and the instrument will return to
the RUN mode.
NOTICE
Before performing a calibration,
•
the transducer must be connected to the instrument,
•
the channel’s Excitation and Signal Jumpers must be set,
•
the
CALIBRATION TYPE
must be selected (see “
on page 106), and
•
the
CALIBRATION DATA
must be entered (see “
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.
Summary of Contents for SC500
Page 8: ...page 8 008 0671 00 ...
Page 14: ...page 14 008 0671 00 ...
Page 18: ...page 18 008 0671 00 ...
Page 24: ...page 24 008 0671 00 ...
Page 38: ...page 38 008 0671 00 ...
Page 42: ...page 42 008 0671 00 ...
Page 50: ...page 50 008 0671 00 ...
Page 72: ...page 72 008 0671 00 ...
Page 82: ...page 82 008 0671 00 ...
Page 90: ...page 90 008 0671 00 ...
Page 116: ...page 116 008 0671 00 ...
Page 126: ...page 126 008 0671 00 ...